Key points
From the paper's conclusion- Background The PRAETORIAN score (PS) estimates the risk of defibrillation failure after subcutaneous implantable cardioverter‐defibrillator (S‐ICD) implantation based on postoperative chest radiography in the standing position.
Abstract
BACKGROUND: The PRAETORIAN score (PS) estimates the risk of defibrillation failure after subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation based on postoperative chest radiography in the standing position. However, the usefulness of assessing the PS in the supine position intraoperatively has not yet been established. METHODS: We evaluated conventional (standing) and intraoperative supine PS in 124 consecutive Japanese patients undergoing S-ICD implantation. Changes between conventional and supine PS values were analyzed in two ways: (1) the presence of any difference in the absolute PS values regardless of the risk category, and (2) changes in clinically relevant risk categories for conversion failure (low, intermediate, or high risk). Predictors of discordance between the two assessments were also investigated. RESULTS: The median conventional PS was 30; 116 and 8 patients were classified as having low and intermediate risk of conversion failure, respectively. Numerical differences between the conventional and supine PS values were observed in 13 patients. However, clinically meaningful risk reclassification occurred in only two patients, suggesting that intraoperative supine PS assessment may underestimate PRAETORIAN score-based risk classification in a small subset of patients. Among patients with numerical PS changes, supine PS was significantly lower than conventional PS. An elevated BMI was strongly associated with PS discordance. Exploratory receiver operating characteristic curve analysis identified a BMI cutoff of 25.6 kg/m2 for discriminating patients with numerical PS discordance. CONCLUSIONS: Intraoperative supine PS assessment shows excellent concordance with conventional standing PS and may provide a practical tool for confirming appropriate system positioning in most Japanese patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Subcutaneous versus transvenous implantable cardioverter-defibrillator therapy: an updated systematic review and meta-analysis.Open heart · 2026
- Implantation Technique, Continuous Implant Geometry, and Longitudinal Low-Voltage Impedance After Subcutaneous Implantable Cardioverter-Defibrillator Implantation.Journal of cardiovascular electrophysiology · 2026
- Clinical outcomes and cost-effectiveness of subpectoral implantation as salvage strategy for cardiac implantable electronic device pocket infection: A comparative retrospective study.Journal of plastic, reconstructive & aesthetic surgery : JPRAS · 2026
- Septal-Inferior Vena Cava Angle and Accurate Septal Implantation of Leadless Pacemakers: Computed Tomography-Based Study.Pacing and clinical electrophysiology : PACE · 2026
- Late Obesity-Driven Subcutaneous ICD Geometric Failure: Salvage With Sub-Serratus Generator Repositioning.Pacing and clinical electrophysiology : PACE · 2026
- Impact of anatomical factors and implantation position on defibrillation threshold in extravascular implantable cardioverter-defibrillatorEuropean Heart Journal · 2026