5.0 T Coronary Magnetic Resonance Angiography in Calcified Coronary Arteries: Diagnostic Performance Compared With Computed Tomography Using Invasive Angiography as Reference.
Tingyi Du, Lan Lan, Zhe Dong, Gonghao Ling, Rongqing Sun, Yuan Zhou, Haibo Xu
Department of Radiology, Zhongnan Hospital of Wuhan University, Wuhan, Hubei, China. · Department of Cardiovascular Surgery, Zhongnan Hospital of Wuhan University, Wuhan, Hubei, China.
Key points
From the paper's conclusion- The performance of 5.0 T Coronary MR angiography (CMRA) in calcified arteries remains unclear.
Abstract
BACKGROUND: Coronary computed tomography angiography (CCTA) often overestimates stenosis in calcified plaques due to blooming artifacts. The performance of 5.0 T Coronary MR angiography (CMRA) in calcified arteries remains unclear. PURPOSE: To assess plaque effects on 5.0 T CMRA image quality and compare diagnostic performance with CCTA. STUDY TYPE: Prospective. POPULATION: Forty-six patients with suspected coronary artery disease. FIELD STRENGTH/SEQUENCE: 5.0 T; free-breathing, non-contrast 3D gradient-echo sequence, Dixon water-fat separation, artificial intelligence-assisted compressed sensing. ASSESSMENT: Three blinded radiologists with 6, 8, and 12 years' experience independently evaluated all datasets. Obstructive stenosis was defined as ≥ 50%; CMRA and CCTA image quality used a 4-point scale; quantitative CMRA measures included vessel edge sharpness (VES), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR). CCTA assessed plaque composition, calcium burden, and calcification length. Invasive angiography was the reference. STATISTICAL TESTS: Diagnostic metrics and generalized estimating equations evaluated performance and plaque-image-quality associations. Wilcoxon tests compared visual scores. Agreements used Cohen's κ, Fleiss' κ, and intraclass correlation coefficients. RESULTS: 5.0 T CMRA showed higher specificity (90.1% vs. 77.7%, p < 0.001) and accuracy (89.4% vs. 82.8%, p = 0.006) than CCTA, with similar sensitivity (88.7% vs. 87.7%, p = 0.740). For CMRA, VES was lower in segments containing mixed plaques than in noncalcified or calcified plaques (16.0 ± 0.8 vs. 18.2 ± 0.8 and 18.3 ± 1.0; p < 0.001). Visual scores differed across calcium-burden categories (3.35 ± 0.07, 3.16 ± 0.09, 2.96 ± 0.09, and 2.79 ± 0.09; p < 0.001). Greater calcification length was associated with lower VES (B = -0.152, p = 0.008) and visual scores (B = -0.019, p = 0.019). DATA CONCLUSION: In this cohort, 5.0 T CMRA showed higher specificity and accuracy than CCTA, with similar sensitivity. Plaque characteristics were associated with selected CMRA image-quality measures. EVIDENCE LEVEL: 2. TECHNICAL EFFICACY: Stage 2.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessment of Coronary Plaque Burden using Non-Contrast T1-weighted Cardiovascular MR: Comparison with Coronary CT Angiography.Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance · 2026
- Detection of calcified plaques: comparison between coronary CT angiography and thin-slice non-contrast CT with deep learning-aided image registrationEuropean Radiology · 2026
- Simulation of calcium blooming artifacts in coronary CT angiography from digital phantoms.Cardiovascular diagnosis and therapy · 2026
- Evaluation of Coronary Artery Disease Plaques Using Coronary CT Angiography in a High-Risk Population.Cureus · 2026
- Combined non-contrast enhancement 3D bright-blood and 2D black-blood coronary magnetic resonance angiography for the assessment of coronary artery stenosis.Quantitative imaging in medicine and surgery · 2026
- Deep learning-based assessment of coronary artery disease using curved-multiplanar reconstruction: comprehensive evaluation of stenosis, calcification, and plaqueEuropean Heart Journal · 2026