Effect of Driver Material Properties and Excitation Frequencies on Quantitative Myocardial Stiffness Measurements in Magnetic Resonance Elastography
Yong Cheng, Jun Wang, Shuangshuang Xu, Zijun Zheng, Yewen Jiang, Peiyang Zheng, Shutian An, Zhanglei Li,
Key points
From the paper's conclusion- Appropriate matching of driver materials and frequencies may help optimize wave field quality, thereby providing more reliable myocardial stiffness estimates.
Abstract
Background Myocardial stiffness is a crucial indicator of cardiac function and is associated with various cardiovascular diseases. Magnetic resonance elastography (MRE) is a noninvasive imaging technique used to measure myocardial stiffness. However, the effects of driver material properties and excitation frequencies on MRE performance remain unclear. This study investigated the influence of different driver materials and excitation frequencies on myocardial stiffness measurements and signal quality. Methods This prospective study included 13 healthy adults (four women and nine men) without known cardiac disease. MRE was performed on a 3.0‐T magnetic resonance imaging system using two different driver materials (flexible and rigid) at two frequencies (80 and 140 Hz). Myocardial stiffness and octahedral shear strain signal‐to‐noise ratio (OSS‐SNR) were quantified. Paired t ‐test, intraclass correlation coefficient, and Bland–Altman analysis were used for statistical analysis. Results At 80 Hz, the flexible driver yielded significantly higher stiffness values and OSS‐SNR than the rigid driver (all p < 0.01). At 140 Hz, the two drivers resulted in similar stiffness values; however, the rigid driver yielded a significantly higher OSS‐SNR ( p = 0.007). A significant increase in stiffness was observed at 140 Hz compared with that at 80 Hz ( p < 0.001). Conclusions Owing to its superior signal quality, the flexible driver may be more suitable for low‐frequency (80 Hz) MRE. Meanwhile, the rigid driver may exhibit better performance at higher frequencies (140 Hz). Appropriate matching of driver materials and frequencies may help optimize wave field quality, thereby providing more reliable myocardial stiffness estimates.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating the impact of passive driver position on image consistency and repeatability in 3.0T cardiac magnetic resonance elastography.Frontiers in radiology · 2026
- Reproducibility of 3.0 T single-frequency cardiac MR elastography for quantification of left ventricular myocardial stiffness.Radiology advances · 2026
- Brain mechanics and cognition: the role of rodent MR elastography in identifying cellular mechanisms.Frontiers in cognition · 2026
- Multifrequency neural network-based wave inversion in MR elastography with uncertainty quantification.Medical image analysis · 2026
- Chronic Liver Disease: Assessing Inflammation and Fibrosis Using Three‐Dimensional MR Elastography With Same‐Day Biopsy in a Prospective CohortJournal of Magnetic Resonance Imaging · 2026
- Magnetic Resonance Elastography of Regional Aortic Mechanics and True- and False-Lumen Stiffness in Aortic Dissection: A Feasibility Study.medRxiv : the preprint server for health sciences · 2026