On-Column Focusing Strategies for Large-Volume Injections in Nano-Liquid Chromatography: Sample Diluent, Injection Sequence, and Localized Cooling.
Giovanni D'Orazio, Salvatore Fanali
Istituto di Biologia e Biotecnologia Agraria, CNR-Consiglio Nazionale delle Ricerche, Montelibretti, Rome, Italy. · Institute of Physical and Analytical Chemistry, School of Exact and Natural Sciences, Tbilisi State University, Tbilisi, Georgia.
Key points
From the paper's conclusion- These results demonstrate that combining diluent tuning, injection-sequence control, and low-temperature focusing can extend nano-LC injection volumes by nearly two orders of magnitude beyond the theoretical admissible injection volume while preserving separation performance, offering simple, low-cost strategies to improve analytical sensitivity.
Abstract
Sensitivity remains a major limitation of nano-liquid chromatography (nano-LC), where reduced injection volumes and small spectrophotometric detection cells constrain analytical performance relative to conventional HPLC. This work systematically investigated sample on-column focusing strategies to control overloading during large-volume injections in nano-LC without compromising chromatographic resolution and efficiency. Using a model mixture of alkylbenzenes separated on a 100 µm i.d. capillary column packed with a C18 stationary phase in a lab-made nano-LC system, three complementary approaches were evaluated: (i) tuning the sample diluent composition, including aqueous content and the use of conventional and green organic co-solvents (isoamylacetate, 2-methyltetrahydrofuran, and dimethyl carbonate); (ii) injection-sequence strategies based on pre- and post-injection solvent plugs; and (iii) localized low-temperature focusing achieved with an instant cold spray applied at the column inlet immediately before sample injection. Optimizing diluent composition alone enabled a ∼40/60-fold increase in injection volume (up to 600-1000 nL) without loss of efficiency. Adding a post-injection water plug extended this to ∼120/150-fold (2000-2500 nL), improving chromatographic efficiency by up to 30%. When high diluent eluent strength precluded solvent-based focusing, a brief cold-spray application at the column inlet (down to ∼4°C) fully restored resolution, independent of the organic solvent used. These results demonstrate that combining diluent tuning, injection-sequence control, and low-temperature focusing can extend nano-LC injection volumes by nearly two orders of magnitude beyond the theoretical admissible injection volume while preserving separation performance, offering simple, low-cost strategies to improve analytical sensitivity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A facile in-loop dilution strategy for large volume injection in ultra-sensitive UPLC-MS/MS determination of 37 veterinary drug residues in beef.Talanta · 2026
- Trap-Refocusing-Elute NanoLC-MS: A Robust Strategy Combining Trap-and-Elute Configuration with Direct-Injection Chromatographic Performance.Analytical chemistry · 2026
- Optimizing nanoLC-TIMS-TOF and lipidomics sample preparation for low volume of plasmaScientific Reports · 2026
- Continuous sampling/transfer strategy to overcome sampling-volume mismatch in heart-cutting two-dimensional liquid chromatography: Application to ertapenem impurity profiling.Journal of chromatography. A · 2026
- Post-Column Dead Volume as a Critical Determinant in Low-Input, High-Throughput Proteomics.Analytical chemistry · 2026
- Optimization of Electrospray-Scanning Mobility Particle Sizer (ES-SMPS) Residual Control and Comparative Evaluation of Low-Concentration Silica Nanoparticle Characterization.Langmuir : the ACS journal of surfaces and colloids · 2026