Drug testing and analysis

Diclofenac in Urine Samples Deactivates the GC/C/IRMS Reactor and Compromises the Determination of δ13C Values of Steroids.

Yan Mengmeng, Wen Chao, Yan Kuan, Zhu Tianshuo, Ma Congcong, Wang Zhanliang

Beijing Anti-Doping Laboratory, Beijing Sport University, Beijing, China.

Key points

From the paper's conclusion
  • During subsequent GC/C/IRMS analysis, residual chlorinated substances in purified fractions may contribute to oxidation reactor deactivation, thereby interfering with the determination of analyte 's δ13C values.

Abstract

Gas chromatography/combustion/isotope ratio mass spectrometry (GC/C/IRMS) is used as a confirmation procedure for detecting the administration of synthetic forms of endogenous anabolic androgenic steroids (EAAS) by comparing the carbon isotope ratios (δ13C) of target compounds (TCs) with those of endogenous reference compounds (ERCs). In recent years, some samples have been found to still contain interfering matrix components in different TC/ERC fractions after HPLC purification, which can rapidly deactivate the GC/C/IRMS reactor. 1-(2,6-Dichlorophenyl)indolin-2-one (1-(DCP)I) has been identified as a major contributor to reactor deactivation. Subsequent studies suggested that 1-(DCP)I was associated with diclofenac (DCF), a major metabolite formed following DCF sodium administration. DCF can persist in target fractions even after purification by one- or two-dimensional semipreparative liquid chromatography coupled with fraction collection. During subsequent GC/C/IRMS analysis, residual chlorinated substances in purified fractions may contribute to oxidation reactor deactivation, thereby interfering with the determination of analyte 's δ13C values. Optimization of the backflush time in the GC/C/IRMS system can effectively minimize the influence of these chlorine-containing substances on analysis.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Diclofenac in Urine Samples Deactivates the GC/C/IRMS Reactor and Compromises the Determination of δ13C Values of Steroids — Science Skim