Effect of Wall Motion Sampling on CFD-Derived Left Atrial Flow Metrics.
Yvonne Stöcker, Manuel Guerrero-Hurtado, Eduardo Durán, Alejandro Gonzalo, Zorica Ristić, Åshild Telle, Ahmad Kassar, Romanos Haykal,
Institute of Fluid Mechanics and Heat Transfer, TU Wien, Vienna, Austria. · Hospital General Universitario Gregorio Marañón, Madrid, Spain.
Key points
From the paper's conclusion- Therefore, temporal interpolation to reconstruct left atrial (LA) wall motion and boundary conditions is required, but its impact on hemodynamic predictions has not been systematically characterized.
Abstract
The temporal resolution of medical imaging sequences used to drive patient-specific computational fluid dynamics (CFD) simulations remains limited, typically providing 10-20 frames per cardiac cycle. Therefore, temporal interpolation to reconstruct left atrial (LA) wall motion and boundary conditions is required, but its impact on hemodynamic predictions has not been systematically characterized. To investigate this, we constructed high-temporal-resolution reference wall-motion data using electromechanical (EM) simulations on five patient-specific atrial geometries with a history of atrial fibrillation. We then generated temporally downsampled datasets to emulate clinical frame rates (5, 10, 20, and 40 frames per cycle) and performed CFD simulations to isolate the effects of temporal undersampling on hemodynamic metrics. The focus was placed on kinetic energy, KE, and residence time, T R , particularly in the left atrial appendage (LAA), where thrombosis is most likely to occur. We employed an immersed boundary method to prescribe the wall motion and computed blood T R through a passive scalar transport equation. Results indicate that while global LA hemodynamic indices were marginally affected by the frame rate (errors < 9 % ), LAA metrics were more sensitive with errors up to 31% compared to reference values. The results based on 20 and 40 frames per cycle yielded favorable agreement with reference results, while 5- and 10-frame reconstructions showed larger, though not systematically biased, deviations from the reference. Importantly, patient ranking by blood-stasis indices was largely preserved. The analysis suggests that patient-specific LA reconstructions derived from dynamic CT imaging provide a reliable basis for estimating LAA blood-stasis indices. Higher frame rates ( ≥ 20 per cycle) offer improved quantitative accuracy, while lower temporal resolutions may remain informative for patient stratification purposes, where relative ranking is more relevant than absolute accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Volume-Constrained Wall Motion Alters Left Atrial Appendage Hemodynamic Descriptors: A CFD Sensitivity Study.Medical engineering & physics · 2026
- Fibrosis burden and spatial pattern affect left atrial appendage stasis in multi-beat multi-physics simulationsbioRxiv · 2026
- Fluid exchange and renewal in the left atrial appendagearXiv · 2026
- Soft robotic and computational models of left atrial hemodynamics in sinus rhythm and atrial fibrillationNature Communications · 2026
- Quantifying the Functional Effect of Transmural Interpolation Strategies on Atrial Wave Propagation Using a Custom Open-Source Pipeline.Journal of visualized experiments : JoVE · 2026
- Disentangling the Hemodynamic Effects of A-Wave Loss and Cardiac Output Reduction in Atrial Fibrillation.Annals of biomedical engineering · 2026