Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

Comparative Outcomes of Robotic-Assisted Versus Manual Percutaneous Coronary Intervention Using a Four-Channel Robotic Platform.

Junghoon Lee, Daeung Ohn, Sook Jung Kim, Yeon-Jik Choi, Osung Kwon, Jeong-Eun Yi, Sung-Won Jang, Suk-Min Seo

Department of Internal Medicine, Division of Cardiology, Eunpyeong St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.

Trial · n=40

Key points

From the paper's conclusion
  • R-PCI with a four-channel platform was feasible and safe, with lower operator radiation, lower workload, and lower per-minute patient exposure, and allowed remote dual-wire side-branch protection.

Abstract

BACKGROUND: Robotic-assisted PCI (R-PCI) reduces operator radiation and physical workload. Existing platforms operate a single guidewire at a time, and comparative data on R-PCI in situations that require a second wire for side-branch protection are lacking. AIMS: We evaluated the feasibility and safety of R-PCI using a four-channel platform, compared with manual PCI (M-PCI), in elective single-vessel PCI. METHODS: In this single-center pilot study, 40 patients were enrolled (M-PCI, n = 30; R-PCI, n = 10). The four-channel architecture enabled remote dual-wire side-branch protection in 3 of the 10 R-PCI cases. Procedure time, operator and patient radiation, contrast volume, NASA Task Load Index (NASA-TLX), and CK-MB and troponin-I at 6 and 12 h were recorded. Pre-specified per-minute rates were calculated. Groups were compared with the Mann-Whitney U test. RESULTS: Technical and clinical success were 100%, with no conversion to manual operation and no device-related adverse events. R-PCI procedures were longer than M-PCI (33.3 ± 11.4 vs. 26.0 ± 15.1 min; p = 0.027). Operator radiation outside the lead apron was 92% lower with R-PCI (p < 0.001). Cumulative patient radiation and contrast did not differ significantly, but per-minute rates were lower for contrast (-43%; p < 0.001), DAP (-35%; p = 0.002), and air kerma (-42%; p < 0.001). Operator workload (NASA-TLX) was 68% lower (p = 0.002). Post-procedural CK-MB was lower at 6 and 12 h (p = 0.049 and 0.044), and troponin-I followed the same direction without reaching significance. CONCLUSIONS: R-PCI with a four-channel platform was feasible and safe, with lower operator radiation, lower workload, and lower per-minute patient exposure, and allowed remote dual-wire side-branch protection. CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov identifier: NCT06735716 (LNR-AVIARMX2-I02). Registered prior to first patient enrollment.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Comparative Outcomes of Robotic-Assisted Versus Manual Percutaneous Coronary Intervention Using a Four-Channel Robotic Platform — Science Skim