Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

A Contemporary Analysis of Microaxial Pump-Associated Cardiac Perforation Events From the FDA MAUDE Database.

Soomal Rafique, Andrew Sagalov, Ra'Ad Al Tamimi, Fares Ghanem, Raj Patel, Navin Kapur, Abdul Moiz Hafiz

SIU School of Medicine, Springfield, Illinois, USA. · Montefiore Medical Center, Bronx, New York State, USA.

Key points

From the paper's conclusion
  • LV perforation reported with Impella devices is rare but carries high mortality; earlier recognition and prompt repair are associated with improved survival.

Abstract

BACKGROUND: Percutaneous microaxial pumps are used in cardiogenic shock and high-risk percutaneous coronary intervention (PCI), but left ventricular (LV) perforation is an infrequent but serious complication. In 2023, an FDA advisory warned of increased perforation risk. AIMS: To describe clinical setting, patterns, high-risk subgroups, and outcomes of LV perforation in real-world reports. METHODS: We queried the FDA Manufacturer and User Facility Device Experience (MAUDE) database (2016-2024) for Impella adverse-event reports, screened narratives for perforation terms, and manually adjudicated suspected cases to confirm LV perforation and extract device, clinical setting, presentation, management, and outcomes. RESULTS: Among 24,424 Impella reports, 188 involved LV perforation (0.77%). Impella CP accounted for 52.7% and Impella 5.5 for 41.5% (few with 5.0/2.5; none with RP). Annual LV perforations per 1000 reports declined from 166.7 (2016) to 5.2 (2024). Events occurred most often during cardiac surgery (27.1%), acute coronary syndrome (19.1%), or high-risk PCI (17.0%). Intraprocedural diagnosis occurred in 76.2% and was more common with 5.5 versus CP (83.3% vs. 67.7%). With intraprocedural diagnosis, mortality was 24.6% for 5.5 and 28.4% for CP, versus 66.7% for both when diagnosis was post-procedural. Pericardial effusion (34.9%) and hypotension (28.6%) were common. Repair was attempted at 72.9% and reduced mortality, particularly for CP (19.7% with repair vs. 64.7% without; p < 0.001). Overall mortality was 30.2%, without differences by sex or race. CONCLUSIONS: LV perforation reported with Impella devices is rare but carries high mortality; earlier recognition and prompt repair are associated with improved survival.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A Contemporary Analysis of Microaxial Pump-Associated Cardiac Perforation Events From the FDA MAUDE Database — Science Skim