A Contemporary Analysis of Microaxial Pump-Associated Cardiac Perforation Events From the FDA MAUDE Database.
Soomal Rafique, Andrew Sagalov, Ra'Ad Al Tamimi, Fares Ghanem, Raj Patel, Navin Kapur, Abdul Moiz Hafiz
SIU School of Medicine, Springfield, Illinois, USA. · Montefiore Medical Center, Bronx, New York State, USA.
Key points
From the paper's conclusion- LV perforation reported with Impella devices is rare but carries high mortality; earlier recognition and prompt repair are associated with improved survival.
Abstract
BACKGROUND: Percutaneous microaxial pumps are used in cardiogenic shock and high-risk percutaneous coronary intervention (PCI), but left ventricular (LV) perforation is an infrequent but serious complication. In 2023, an FDA advisory warned of increased perforation risk. AIMS: To describe clinical setting, patterns, high-risk subgroups, and outcomes of LV perforation in real-world reports. METHODS: We queried the FDA Manufacturer and User Facility Device Experience (MAUDE) database (2016-2024) for Impella adverse-event reports, screened narratives for perforation terms, and manually adjudicated suspected cases to confirm LV perforation and extract device, clinical setting, presentation, management, and outcomes. RESULTS: Among 24,424 Impella reports, 188 involved LV perforation (0.77%). Impella CP accounted for 52.7% and Impella 5.5 for 41.5% (few with 5.0/2.5; none with RP). Annual LV perforations per 1000 reports declined from 166.7 (2016) to 5.2 (2024). Events occurred most often during cardiac surgery (27.1%), acute coronary syndrome (19.1%), or high-risk PCI (17.0%). Intraprocedural diagnosis occurred in 76.2% and was more common with 5.5 versus CP (83.3% vs. 67.7%). With intraprocedural diagnosis, mortality was 24.6% for 5.5 and 28.4% for CP, versus 66.7% for both when diagnosis was post-procedural. Pericardial effusion (34.9%) and hypotension (28.6%) were common. Repair was attempted at 72.9% and reduced mortality, particularly for CP (19.7% with repair vs. 64.7% without; p < 0.001). Overall mortality was 30.2%, without differences by sex or race. CONCLUSIONS: LV perforation reported with Impella devices is rare but carries high mortality; earlier recognition and prompt repair are associated with improved survival.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Bleeding and thrombotic events during micro-axial flow pump support: A European multicenter cohort studyCritical Care · 2026
- Routine microaxial left ventricular support during nonemergent high-risk PCI: an updated systematic review and meta-analysis with Bayesian and grade methodologies.Expert review of medical devices · 2026
- Short- and intermediate-term outcomes of microaxial flow pump-assisted versus standard-care PCI in high-risk patients: A propensity score-matched analysis.Cardiovascular revascularization medicine : including molecular interventions · 2026
- Outcomes of microaxial flow pump mechanical circulatory support in cardiogenic shockEuropean Heart Journal · 2026
- Radiologic predictors of access-site complications during support with micro-axial flow pump support in infarct-related cardiogenic shock and high-risk percutaneous coronary interventionEuropean Heart Journal · 2026
- Timing of Microaxial Flow Pump Support and In-Hospital Mortality in Infarct-Related Cardiogenic Shock: The TIMING-CS Study.Journal of cardiovascular development and disease · 2026