Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

Intravascular Ultrasound Catheter Entrapment Caused by Extensive Twisting in the Left Main Coronary Artery: A Challenging Retrieval.

Takashi Makinouchi, Tatsuro Yamazaki, Satoshi Tokimasa, Tatsuhiko Asano

Department of Cardiology, Chiba Rosai Hospital, Ichihara, Chiba, Japan. · Department of Cardiovascular Medicine, Chiba Rosai Hospital, Ichihara, Chiba, Japan.

Key points

From the paper's conclusion
  • We report a rare case of intravascular ultrasound (IVUS) catheter entrapment in which extensive twisting of the catheter shaft prevented retrieval.

Abstract

We report a rare case of intravascular ultrasound (IVUS) catheter entrapment in which extensive twisting of the catheter shaft prevented retrieval. In this phenomenon, termed "catheter suicide," the mechanical IVUS catheter twisted within the left main coronary artery (LMCA) and formed complex loops. Initial removal attempts were unsuccessful, and, during these bailout attempts, the imaging core of the IVUS catheter was found to be severed in its proximal portion. Subsequent imaging of another vessel confirmed extensive twisting, with associated iatrogenic dissection, as the cause of entrapment. Ex vivo experiments reproduced this rotational torque-driven complication. The twist was eventually released incidentally when the IVUS catheter was advanced further, and a larger guiding catheter was used, thereby facilitating safe retrieval. Subsequently, a stent was placed to treat the LMCA dissection, achieving thrombolysis in myocardial infarction (TIMI) grade 3 flow, and the patient was discharged without complications. This case illustrates a severe complication specific to mechanical IVUS systems and emphasizes the importance of prevention and early surgical consultation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.