Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions

Mitral Transcatheter Edge-to-Edge Repair at Centers With and Without On-Site Cardiac Surgery: A Comparative Registry With a Shared Interventional Team.

Giulia Marchisio, Elisa Pelloni, Francesco Pisano, Lorenzo Zaccaro, Laura Contu, Maurizio Tusa, Nedy Brambilla, Marco Guerrini,

Division of Cardiology, Umberto Parini Regional Hospital, Azienda USL della Valle d'Aosta, Aosta, Italy. · Clinical and Interventional Cardiology Department, IRCCS Policlinico San Donato, San Donato Milanese, Milan, Italy.

Observational · n=64

Key points

From the paper's conclusion
  • In this preliminary same-team comparison, 30-day MVARC device and procedural success rates were 81.0% at the no-CCH center and 72.1% at the CCH center. Confidence intervals were wide and non-inferiority was not established. These findings support prospective evaluation of network-based M-TEER programs with shared interventional expertise and formal surgical backup.

Abstract

BACKGROUND: Comparative evidence for M-TEER performed at centers with versus without on-site cardiac surgery is lacking. We compared early outcomes at two centers that shared the same interventional operator team but differed in the availability of on-site cardiac surgery. METHODS: We retrospectively analyzed prospectively collected registry data from 64 consecutive patients with severe (4+) MR treated at a center without on-site cardiac surgery (no-CCH, n = 21) and a center with on-site cardiac surgery (CCH, n = 43). The same interventional operator team performed every procedure. At the no-CCH center, surgical expertise was provided through a shared Heart Team and a formal emergency transfer pathway. The cohorts were treated in sequential calendar periods. Principal comparative endpoints were MVARC device and procedural success at 30 days. Exploratory non-inferiority analyses used margins of -12.5% and -10%. RESULTS: MVARC technical success was 100% versus 97.7%. Device and procedural success were each 81.0% versus 72.1% (risk difference [RD] +8.9%, 95% CI -14.9 to +27.5; p = 0.547). Freedom from major MVARC serious adverse events was 95.2% versus 95.3%. Twelve of 16 device-success failures involved a mean transmitral gradient ≥ 5 mmHg. Surgical conversion and in-hospital death were 0% in both centers. CONCLUSIONS: In this preliminary same-team comparison, 30-day MVARC device and procedural success rates were 81.0% at the no-CCH center and 72.1% at the CCH center. Confidence intervals were wide and non-inferiority was not established. These findings support prospective evaluation of network-based M-TEER programs with shared interventional expertise and formal surgical backup.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Mitral Transcatheter Edge-to-Edge Repair at Centers With and Without On-Site Cardiac Surgery: A Comparative Registry With a Shared Interventional Team — Science Skim