Key points
From the paper's conclusion- Diagnostic accuracy for detection of AD pathology was 93% (95% confidence interval 91-95) with 23% intermediate results. Reduced estimated glomerular filtration rate was associated with increased intermediate results among amyloid-negative individuals. Result frequencies remained stable, average imprecision was 8%, and analytical bias between lots and version changes were ≤ 11%.
Abstract
INTRODUCTION: Plasma phosphorylated tau 217 (p‑tau217) immunoassays are increasingly being used to assess Alzheimer's disease (AD) pathology despite limited data affirming test reliability over time. METHODS: Clinical performance of the Fujirebio Lumipulse G p‑tau217 assay was assessed in 744 Mayo Clinic patients presenting for evaluation of cognitive impairment between July 2024 and December 2025. Cerebrospinal fluid p‑tau181/amyloid beta 42 or amyloid positron emission tomography, within 1 year of p-tau217 testing, were used as reference. Analytical robustness was evaluated by tracking result frequencies over time, conducting quality control checks, performing 10 lot‑to‑lot comparisons, and comparing two versions of reagents. RESULTS: Diagnostic accuracy for detection of AD pathology was 93% (95% confidence interval 91-95) with 23% intermediate results. Reduced estimated glomerular filtration rate was associated with increased intermediate results among amyloid-negative individuals. Result frequencies remained stable, average imprecision was 8%, and analytical bias between lots and version changes were ≤ 11%. DISCUSSION: The findings support plasma p-tau217 reliability in clinical practice for evaluation of cognitively impaired individuals.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multi-platform comparison of plasma phosphorylated tau assays: analytical and workflow factors drive clinical implementation decisions.Alzheimer's & dementia (Amsterdam, Netherlands) · 2026
- Clinical performance and decision thresholds of an automated plasma p-tau217 assay for amyloid pathology in a Taiwanese memory clinic.Journal of neurology · 2026
- Analytical validation and amyloid-status discrimination of a high-throughput, research-use-only plasma p-Tau217 immunoassaymedRxiv · 2026
- pTau217 immunoassays in Alzheimer's disease continuum: A comparative study of Ella and Lumipulse.Journal of Alzheimer's disease : JAD · 2026
- Implementation and clinical concordance of a plasma pTau217 RUO assay within an in-house IVD/LDT workflowJournal of Laboratory Medicine · 2026
- Clinical utility of LUMIPULSE plasma p-tau217 to identify amyloid positivity among candidates for amyloid-targeting treatments in a real-world cohort.Alzheimer's & dementia (Amsterdam, Netherlands) · 2026