A Stretchable, Sensitive, and Stable Cluster‐Hydrogel Electrode for Long‐Term Clinical Surveillance of Parkinson's Disease
Yuqin Zhang, Yufeng Ke, Shuangjie Liu, Yang Wang, Leyi Jia, Xinyu Sun, Shaoya Yin, Zhuo Chen,
Tianjin University · Tianjin Huanhu Hospital
Key points
From the paper's conclusion- The resulting cluster‐hydrogel (CH) electrode exhibits an ultralow contact impedance of 6.9 kΩ·cm 2 , equivalent to 1/6 th that of conventional Ag/AgCl electrodes and 1/24 th that of metallic electrodes, while sustaining a high signal‐to‐noise ratio of 6.6 dB over 12 h continuous recording.
Abstract
ABSTRACT Neural recording electrodes are pivotal components of a brain‐computer interface (BCI) for diagnosing neurological diseases. However, the conventional electroencephalography (EEG) electrodes face challenges in long‐term stability and biocompatibility. The contemporary dry electrodes offer enhanced stability but remain suboptimal in recording sensitivity. To address these constraints, we developed mechanically flexible, high‐sensitivity, biocompatible electrodes employing a redox‐active multienzyme‐mimetic gold cluster integrated into highly conductive polymer hydrogel networks for prolonged neural monitoring in Parkinson's disease (PD). The resulting cluster‐hydrogel (CH) electrode exhibits an ultralow contact impedance of 6.9 kΩ·cm 2 , equivalent to 1/6 th that of conventional Ag/AgCl electrodes and 1/24 th that of metallic electrodes, while sustaining a high signal‐to‐noise ratio of 6.6 dB over 12 h continuous recording. Clinically, through the integration of machine learning techniques, the CH electrodes enabled accurate PD diagnosis via alpha reactivity and entropy analyses, achieving an area under the curve (AUC) of 0.9 (KNN classifier) and a phase‐locking value (PLV) Pearson correlation of 0.934. In addition, the CH electrodes outperformed conventional EEG electrodes in alpha‐band sensitivity, achieving a 16.2‐fold higher signal‐to‐noise ratio that enhanced the detection of clinically diagnostic electrophysiological signatures in PD. Combined with high biocompatibility, the CH electrodes hold translational promise as an efficient, safe, and stable neural recording system for clinical BCIs and neuroscience applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multifunctional hydrogel sensors with dynamic covalent networks for machine learning-assisted Parkinson's disease diagnosis and encrypted human-computer interactionMaterials Today Bio · 2025
- Systematic Benchmarking of a Noise‐Tolerant Conductive Hydrogel Electrode for Epidermal BioelectronicsAdvanced Science · 2025
- Skin Conformal Hydrogel Bioelectrodes for High‐Fidelity Electrophysiology and Human–Machine InterfacesAdvanced Healthcare Materials · 2026
- Plasmonic hot-electron-assisted ultra-stretchable hydrogel electrodes for wearable cardiovascular monitoring and AI-driven predictive analyticsJournal of Materials Chemistry B · 2026
- Interference-resistant Janus hydrogel electrodes for AI-enabled platform for rehabilitation assessment.Journal of colloid and interface science · 2026
- Conducting polymers for implantable neural electrodes in brain-computer interfaces.Journal of materials chemistry. B · 2026