Using an instrumented hammer to predict the lateral hinge fracture in high tibial osteotomy: A cadaveric study
Claire Bastard (MSME), Manon Bas-Dit Nugues (MSME), Giuseppe Rosi (MSME), Charles-Henri Flouzat-Lachaniette (MSME), Arnaud Dubory (MSME), Guillaume Haiat (MSME)
Key points
From the paper's conclusion- This cadaveric study demonstrated that the instrumented hammer could prevent impacts leading to hinge rupture, which suggests that future clinical trials are warranted.
Abstract
Open-wedge High-Tibial Osteotomy is the most commonly employed technique addressing varus knee osteoarthritis. During this procedure, the osteotome passes through cortical and trabecular bone. A common complication is lateral hinge fractures, which depends upon when the surgeon stops impacting the osteotome. The aim of the present study is to determine whether an instrumented hammer can be employed to detect (i) the bone type (cortical or trabecular) surrounding the osteotome tip and (ii) when bone is fractured by the osteotome. Seventeen lower-limb specimens were obtained from nine fresh human anatomical subjects. The osteotome was impacted using the instrumented hammer until rupture. Each impact signal was recorded and analyzed. An algorithm was developed to detect the transitions between cortical and trabecular bone and vice versa. Detection was based on the relative variation of $\tau$ derived from the force signal as a function of time as the time difference between the second___and___first impact peaks. The values of $\tau$ were significantly lower in cortical bone compared to trabecular bone. Considering the arrival of the osteotome tip in trabecular bone during a medial, a difference of less than 2 impacts between the surgeon and the algorithm was obtained for 71% of the cases. For the lateral osteotomy it was obtained in 81% of the cases. This cadaveric study demonstrated that the instrumented hammer could prevent impacts leading to hinge rupture, which suggests that future clinical trials are warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- No significant change in tibial torsion despite hinge fracture after high tibial osteotomy using patient-specific instruments.Journal of experimental orthopaedics · 2026
- Hybrid Protective Sentinel System: A Combined Technique for Safe and Controlled Open-Wedge High Tibial Osteotomy.Arthroscopy techniques · 2026
- Biomechanical Effects of Distal Fragment Translation Guided by Target Postoperative Intermetatarsal Angle in Minimally Invasive Transverse Osteotomy for Hallux Valgus: A Patient-Specific Finite Element Study.Bioengineering (Basel, Switzerland) · 2026
- Bioplating demonstrates increasing micromovement during cyclic loading after anterior closed wedge high tibial osteotomy despite no significant difference in failure load compared to angle-stable plate fixation.Journal of experimental orthopaedics · 2026
- Loading conditions during femoral preparation for THA procedures: Mechanical effects of broach holder offset.Injury · 2026
- Detecting the pterygomaxillary disjunction using an instrumented hammer: A cadaveric studyarXiv · 2026