A Compact Single-Flange Low-Energy Inverse Photoemission Spectrometer with Elliptical-Mirror Photon Collection
Jong-Am Hong, Min-Jae Maeng, Kyu-Myung Lee, Seunggu Lee, In-Sung Park, Yongsup Park
Key points
From the paper's conclusion- We compare four photon-collection configurations, each addressing a limitation of its predecessor: a detector beside the sample without collection optics (no-mirror), an off-axis parabolic (OAP) mirror delivering a collimated beam to a head-on photomultiplier tube (PMT), external optics coupling the OAP output to a side-on PMT with about twice the quantum efficiency (OAP-external), and an in-vacuu
Abstract
Low-energy inverse photoemission spectroscopy (LEIPS) is limited by weak photon emission, so photon collection optics largely determine its sensitivity. We compare four photon-collection configurations, each addressing a limitation of its predecessor: a detector beside the sample without collection optics (no-mirror), an off-axis parabolic (OAP) mirror delivering a collimated beam to a head-on photomultiplier tube (PMT), external optics coupling the OAP output to a side-on PMT with about twice the quantum efficiency (OAP-external), and an in-vacuum elliptical mirror focusing the emitted light onto the detector (elliptical). Ray-tracing shows that the elliptical mirror delivers 91.9% of the emitted photons to the detector, compared with 63.3% for the OAP mirror and 3.06% without a mirror, while a plane mirror folding the light path by about 90$^\circ$ allows the electron gun, collection optics and detector to fit on a single 8-inch flange. For C$_{60}$ films, the elliptical configuration gives the highest lowest unoccupied molecular orbital (LUMO) intensity of the four, 271 counts per second per $\mu$A of sample current, 1.74 and 3.23 times those of the OAP and OAP-external configurations and about 45 times that of the no-mirror configuration. The gain over the OAP configuration is comparable to the 1.45-fold increase predicted by ray-tracing, but the filter and detector also differed, so the geometric contribution could not be isolated. The OAP-external configuration gave only 0.54 times the OAP signal despite its higher quantum efficiency. High sensitivity therefore requires both a large collection solid angle and a short optical path kept inside the vacuum.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Sub-100 meV low-energy inverse photoelectron spectroscopy using an electrostatic monochromator.The Review of scientific instruments · 2026
- Multimode objective lens for momentum microscopy and x-ray photoemission electron microscopy: ExperimentsReview of Scientific Instruments · 2026
- Ultraviolet spectrophotometer using photomultiplier tubes in photon counting mode for measurement of optical properties of lithography optics.Optics express · 2026
- Transverse modulation of continuous electron beams by a structured optical cavityJournal of Physics: Photonics · 2026
- Micro- and nanoscale focusing across the extreme ultraviolet range of the ASTRID2 light source with a capillary optic.The Review of scientific instruments · 2026
- Sub-ångström resolution ptychography in a scanning electron microscope at 20 keVNature Communications · 2025