Multi-UTE Myelin Fraction Mapping: Repeatability and Cross-Field-Strength Reproducibility at 1.5T and 3T
Nan Yin, Marco Reisert, Theo Demerath, Shuai Liu, Alexander Rau, Uzay Emir, Horst Urbach, Michael Bock,
Key points
From the paper's conclusion- Purpose: To evaluate the repeatability and cross-field-strength reproducibility (1.5T vs 3T) of in vivo cerebral myelin fraction (MF) mapping with multi-contrast dual-resolution 3D UTE (multi-UTE), and to compare it with magnetization transfer (MT)-based methods.
Abstract
Purpose: To evaluate the repeatability and cross-field-strength reproducibility (1.5T vs 3T) of in vivo cerebral myelin fraction (MF) mapping with multi-contrast dual-resolution 3D UTE (multi-UTE), and to compare it with magnetization transfer (MT)-based methods. Methods: MT methods provide an indirect myelin-sensitive measure derived from magnetization exchange between macromolecular and free-water proton pools, whereas multi-UTE targets the ultrashort-T2* signal components. In five healthy volunteers, multi-UTE and MT-based measurements were performed at 1.5T and 3T, and the calculated MF maps were compared with magnetization transfer ratio (MTR) and magnetization transfer saturation (MTsat) maps. Two volunteers underwent repeated multi-UTE measurements within and between days. Field-strength reproducibility and test-retest repeatability were evaluated using linear regression, Bland-Altman analysis, concordance correlation coefficient (CCC), and intraclass correlation coefficient (ICC). Results: Multi-UTE enabled simultaneous acquisition of quantitative and high-resolution structural images at both 1.5T and 3T. MF repeatability was high at 3T (ICC = 0.94-0.96) and more variable at 1.5T (0,64-0.96). Across field strengths, MF and MTsat showed comparable reproducibility (CCC 0.70 vs 0.60) with small median absolute differences (8.8% vs 8.9%). MTsat showed tighter correlation (R2 = 0.77 vs 0.68), while MF showed a wider inter-regional dynamic range. MTR differed strongly between field strengths (CCC = 0.06). Conclusion: Multi-UTE MF, obtained without B1 or T1 mapping and alongside a high-resolution structural image, achieves cross-field reproducibility comparable to, but not exceeding, MTsat. Lower SNR at 1.5T and unmodelled T1 weighting may have contributed to residual field dependence.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quiet, rapid 3D multiparametric mapping using magnetization-prepared zero echo time MRIarXiv · 2026
- High-Resolution Quantitative MRI Using Contrast-Domain Multiscale Energy (MuSE) RegularizationarXiv · 2026
- Feasibility of a 2-Minute Multi-Echo UTE Acquisition for Simultaneous CT-Like Bone-Weighted Imaging and Quantitative T2* Mapping of Short-T2 TissuemedRxiv · 2026
- Quantitative proton density mapping with data-driven reception profile correction derived from macromolecular proton fraction.medRxiv : the preprint server for health sciences · 2026
- Diffusion Tensor MRI and Spherical‐Deconvolution‐Based Tractography on an Ultra‐Low Field Portable MRI SystemHuman Brain Mapping · 2026
- An Open Test-Retest Dataset of Rapid Myelin Mapping Acquisitions.bioRxiv : the preprint server for biology · 2026