Measurements of the neutron background for the Colorado Underground Research Institute (CURIE)
Dakota K. Keblbeck, Uwe Greife, Eric Mayotte, Wouter Van De Pontseele, Kyle G. Leach
Key points
From the paper's conclusion- We present the first direct measurements of the neutron background at the Colorado Underground Research Institute (CURIE), a shallow-underground facility with an effective overburden of approximately $400$~m.w.e.\ in the Edgar Experimental Mine in Idaho Springs, Colorado.
Abstract
We present the first direct measurements of the neutron background at the Colorado Underground Research Institute (CURIE), a shallow-underground facility with an effective overburden of approximately $400$~m.w.e.\ in the Edgar Experimental Mine in Idaho Springs, Colorado. The radiogenic and muon-induced components were characterized in the Subatomic Particle Hideout. The radiogenic component was measured with a custom Bonner Cylinder Spectrometer comprising a $^{3}$He proportional counter and six cylindrical high-density polyethylene moderators. Its response functions were simulated in \textsc{geant4}, empirically calibrated using a $^{239}$Pu-$^{9}$Be source and the shadow-cone method, and then corrected to account for the approximately isotropic underground field. Unfolding the measured count rates gives a total local scalar neutron flux, dominated by radiogenic production, for $E_n\lesssim\SI{10}{\MeV}$ of $(2.51\pm0.58_{\rm stat}\pm0.39_{\rm sys})\times10^{-5}$~\si{\per\square\centi\meter\per\second}. The muon-induced component was measured with a BC-501A organic liquid scintillator operated for \SI{1431}{\hour}. Pulse shape discrimination identified 55~$\pm$~7 neutron candidates. Using the simulated detector response, these counts yield an integral muon-induced neutron flux at the rock--cavern boundary above \SI{10.5}{\MeV} of $(2.09\pm0.29_{\rm stat}\,{}^{+0.79}_{-0.53}{}_{\rm sys})\times10^{-7}$~\si{\per\square\centi\meter\per\second}. This flux uses the first-entry, boundary-area normalization of a previous CURIE simulation study. Its central value is approximately \SI{12}{\percent} higher than the site-specific prediction but is consistent within the reported uncertainties. Together, these measurements characterize the radiogenic and high-energy cosmogenic neutron environments at CURIE and provide quantitative input for planning future low-background experiments at the facility.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Nuclear and radiological characterization of a 226Ra-Be neutron irradiator at INFISA-UNMSM, Peru, using Monte Carlo simulation with the OpenMC codeNuclear Engineering and Technology · 2026
- Background decomposition of the CONUS+ run 1 dataarXiv · 2026
- Experimental validation of ambient dose equivalent rate calculations for a 241Am-Be neutron source using OpenMC code.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Thick-target yield of 65Cu(α,n)68Ga near threshold and implications for nuclear medicine, deep underground detector backgrounds, and nucleosynthesis.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Estimating cosmogenic backgrounds in nEXO at SNOLAB using Geant4 and FLUKAarXiv · 2026
- Experimental validation of a newly developed neutron ambient dose H* detector in an accelerator-based BNCT fieldJapanese Journal of Applied Physics · 2026