When Magnitude Suffices: Reconstructing Time-Domain-Gated Microwave Signals Without Phase on a Thoracic Phantom
Maciej Cicho\'n, Patrycja Go\l{}a\'s, Maksymilian Bartosik, Angelika Walczak, Agata Bielecka-D\k{a}browa, Maciej \'Slot
Key points
From the paper's conclusion- We show that, on one calibrated measurement configuration, the gated magnitude the VNA would give can be reconstructed from a magnitude-only sweep of a spectrum analyzer with a tracking generator (SA).
Abstract
Time-domain gating separates the arrivals that form a microwave S-parameter measurement. It requires the measurement of complex values, so this application almost always uses a vector network analyzer (VNA). We show that, on one calibrated measurement configuration, the gated magnitude the VNA would give can be reconstructed from a magnitude-only sweep of a spectrum analyzer with a tracking generator (SA). We select the gate center for separability of the measured states, correct each instrument once by a per-bin subtraction of a reference state, and train a fully connected network on sweeps from both instruments. The test object is a torso phantom with lung-shaped compartments whose filling takes eleven dielectric states, each held out in turn. On $S_{11}$ the reconstruction from SA input reaches a shape error of 0.08 dB$^2$ with a state ordering of 0.81. This matches the reconstruction from VNA input. On $S_{21}$ it reaches 0.82 dB$^2$ and keeps the ordered groups of the states. An explicit phase prediction performs much worse, because gating needs the phase to about 0.2 rad RMS. Across eleven measurement configurations the dependence of phase on magnitude is strong for reflection and weak for transmission. This explains the difference between $S_{11}$ and $S_{21}$. Where the magnitude carries enough of the phase, a scalar instrument gives the gated response without any phase measurement after installation. On the open UM-BMID breast data set, with whole phantoms held out, the shape error of the network is 30 percent lower than that of a per-bin linear map.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Observability‐Aware Deep Regression for Single‐Frequency Microwave Damage Localization in Absorbing CoatingsMicrowave and Optical Technology Letters · 2026
- Intelligent Inversion and Calibration Parameter Uncertainty Quantification Method for Microwave Probe S ParametersEngineering Research Express · 2026
- Positioning-robust broadband microwave transmission as a minimal-complexity approach to thoracic dielectric sensing: simulation and phantom analysis.Scientific reports · 2026
- Modification of the method for measuring dielectric and magnetic parameters based on a waveguide transmission line in the microwave regionEngineering Research Express · 2026
- A Prototype Distributed Six-Port Reflectometer-Based Transmission-Only Vector Network Analyzer.Sensors (Basel, Switzerland) · 2026
- End‐to‐end film‐based characterization of gated delivery latencies in a spot‐scanned synchrotron‐based proton therapy systemJournal of Applied Clinical Medical Physics · 2026