Feasibility of a Patient-Specific Digital Twin for Ventriculoperitoneal Shunt Trajectory Planning
Kritsanavis Chongsrid, Vich Yindeedej, Supaporn Konmun, Xulei Yang, Si Yong Yeo
Key points
From the paper's conclusion- We built a digital twin that reconstructs skull, scalp, ear and ventricles from preoperative CT by deterministic geometry, searches every entry around Keen's point for the perpendicular trajectory with the longest intraventricular length under safety constraints, and returns the plan as two ruler measurements from the ear and two depths.
Abstract
Catheter obstruction is the commonest cause of shunt failure; freehand insertion at Keen's point leaves the tip free in cerebrospinal fluid in only one case in ten. We built a digital twin that reconstructs skull, scalp, ear and ventricles from preoperative CT by deterministic geometry, searches every entry around Keen's point for the perpendicular trajectory with the longest intraventricular length under safety constraints, and returns the plan as two ruler measurements from the ear and two depths. Executed freehand in 11 consecutive adults (IDEAL Stage 2a), it achieved Hayhurst Grade 1 placement in 10 of 11 against 22 of 213 historically (p = 9 x 10^-9), median entry and angular deviations 2.5 mm and 5.9 degrees, 1.4 cm more intraventricular catheter than Keen's point, and no intraoperative complication. The technique was unchanged; the gain came from choosing entry and depth per patient. Whether it lowers revision is the next stage's hypothesis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Stereotactic guidance for ventricular catheter placement in ventriculoperitoneal shunt surgery: a state-of-the-art review on technical rationale, clinical outcomes, and future directions.Frontiers in neurology · 2026
- PATTERNS AND MECHANISMS OF VENTRICULOPERITONEAL SHUNT MALFUNCTION ACCORDING TO PROXIMAL VENTRICULAR CATHETER ENTRY POINT IN CHILDREN: A SCOPING REVIEWThe International Journal of Medical Science and Health Research · 2026
- Hybrid anterograde tunnelling for PICC insertion without a dedicated blunt tunneller: A case report.The journal of vascular access · 2026
- Dual-step burr hole with periosteal anchoring technique: a robust method for ventricular catheter fixation - how I do it.Acta neurochirurgica · 2026
- Anatomical and radiological analysis of ventricular catheter trajectories: neuronavigation-guided dissection and tractography-based 3D visualization.Neurosurgical review · 2026
- Mechanical feasibility of retrograde trans-aqueduct access to the third ventricle: a human cadaveric and MRI morphometric study.medRxiv · 2026