Repeatability Characterisation and Error Budget of a Consumer Structured-Light Scanner for 3-D Wound Geometry:A Rigid-Phantom Study
Pushkal Kumar, Aadit Aggarwal, Karlen Aleksanyan
Key points
From the paper's conclusion- We characterise the measurement error of a free- hand consumer structured-light scanner used to derive three- dimensional geometric wound descriptors.
Abstract
We characterise the measurement error of a free- hand consumer structured-light scanner used to derive three- dimensional geometric wound descriptors. Rigid wound-care phantoms cannot change, so every difference between repeat scans of one site is measurement error; all repeats come from a single scanner unit, nine sites and 23 scans, so this charac- terises one instrument. The 95 percent repeatability limit for reconstructed surface area is a factor of 4.8, with a confidence interval from 3.0 to 7.0, so rescanning an unchanged site can shift the reading from a 79 percent decrease to a 377 percent increase. Forty-four of 45 descriptors fall below an intraclass correlation of 0.50, and none of 248 descriptor and pipeline combinations reaches 0.75. An error budget formed by holding the analy- sis region fixed, leaving sensor and reconstruction untouched, bounds the share of variance attributable to how much surface the operator captured at 75 percent for surface area, 91 percent for hull area and 95 percent for bounding-box diagonal; only hull volume is majority instrumental, at 48 percent, so a better sensor would buy little. No wound is delineated anywhere in the chain, so the comparison against the four-week area reduction used clinically to predict healing, a ratio near 2.1, is a lower bound on the noise an unsegmented pipeline must overcome, not a measurement of wound-area reproducibility. Standardising the analysis region cuts the limit to 2.13, meeting that ratio rather than clearing it. Statistical outlier removal imposes a measured systematic area deficit near 11 percent.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluating Neus-Facto Lesion Surface-Area Measurement Using a 3D-Printed Anatomical PhantomApplied Sciences · 2026
- IMTOP, IMage-TO-Print Wound Dressings: From a Single Calibrated RGB Photograph to a Patient-Specific PatchComputers · 2026
- Democratizing three-dimensional surface phenotyping: an open structured-light platform reveals and removes the projection bias in biological imagingbioRxiv · 2026
- Validation of Consumer-Grade 3D Scanners for Clinical Facial Reconstruction: Comparison with a Professional Structured-Light Reference System.Bioengineering (Basel, Switzerland) · 2026
- Comparison of Direct 3D Scans and Plaster Casts in Digital Ankle-Foot Orthosis Workflows.Medical sciences (Basel, Switzerland) · 2026
- Evaluation of Four 3D Facial Scanning Technologies: From Photogrammetry to Structured-Light Systems in Clinical DentistryDentistry Journal · 2026