Balloon Landmark-Assisted Re-entry Using a Re-entry Catheter for Femoropopliteal Chronic Total Occlusion: A Case Report.

Tomohide Sato, Eiji Horio, Kazuhisa Kodama, Tomohiro Sakamoto, Junjiroh Koyama

Division of Cardiology, Saiseikai Kumamoto Hospital, Kumamoto, JPN.

Case report

Key points

From the paper's conclusion
  • We report a case of successful revascularization using an inflated balloon as a fluoroscopic landmark to facilitate re-entry catheter orientation.

Abstract

Re-entry catheters are effective devices for achieving true lumen re-entry during endovascular therapy (EVT) for femoropopliteal chronic total occlusion (FP-CTO). However, accurate orientation of the catheter may be difficult when angiographic visualization of the target lumen is inadequate. We report a case of successful revascularization using an inflated balloon as a fluoroscopic landmark to facilitate re-entry catheter orientation. An 81-year-old man presented with symptomatic left lower extremity ischemia caused by FP-CTO. Two previous EVT procedures had failed because guidewire re-entry into the true lumen could not be achieved after subintimal passage. During the third EVT, the retrograde guidewire remained within the true lumen, whereas the antegrade guidewire advanced within the subintimal space. Because angiographic localization of the re-entry point was difficult, a 5 mm balloon was inflated within the subintimal space and used as a fluoroscopic landmark to determine the orientation of the re-entry catheter. After balloon deflation, the re-entry cannula was advanced from the true lumen toward the subintimal space, enabling successful guidewire passage and subsequent stent implantation. The procedure was completed using only 25 mL of contrast medium. The patient's symptoms resolved, and the ankle-brachial index improved from 0.64 to 1.08. This case suggests that an inflated balloon may serve as a useful fluoroscopic landmark for re-entry catheter orientation when conventional angiographic localization is difficult. In this case, the technique facilitated re-entry catheter orientation and successful guidewire crossing despite limited angiographic visualization. Its potential contribution to contrast minimization, particularly in patients with impaired renal function, requires further evaluation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers