Performance Verification of Four Liver-Related Enzyme Assays on a Dry-Chemistry Biochemistry Analyzer.
Xixi Yuan, Pingping Sun, Qin Jia, Xiaofei Zhang, Linlin Wang, Ji Yang, Zhaojie Huang
Key points
From the paper's conclusion- The verified precision, trueness, linear range, and inter-instrument comparison results for the four analytes fully met the established quality objectives, confirming that the QuidelOrtho VITROS 450 automatic biochemical analyzer is clinically applicable for the detection of these four analytes, thereby ensuring reliable results for routine laboratory testing.
Abstract
BACKGROUND: This study aimed to verify the performance indices of aspartate aminotransferase (AST), alanine aminotransferase (ALT), γ-glutamyl transpeptidase (GGT), and lipase (LPS) and to comprehensively evaluate the performance of the newly commissioned QuidelOrtho VITROS 450 automatic biochemical analyzer in our laboratory, thereby enhancing laboratory work quality. METHODS: The analytical performance of the four items was evaluated via precision, trueness, linearity, and inter-instrument comparison. No unacceptable performance deviations were identified during the evaluation, confirming that the analytical performance of the detection method satisfied the clinical requirements for sample testing. The entire verification process adhered to the referenced standards, including CLSI EP15-A3 and EP09c-Ed3, WS/T 492-2016, WS/T 408-2024 and WS/T 403-2024. RESULTS: The within-run coefficient of variation (CV) of precision verification was less than 1/4 TEa, and the total CV was less than 1/3 TEa. The relative bias of the four items was less than 1/2 total allowable error (TEa). Linearity was verified, and the coefficient of determination (R²) of the four items was ≥ 0.995, which could preliminarily judge that the linear range met the requirements. The comparison between instruments showed that there was no statistically significant difference in the four items between the QuidelOrtho VITROS 450 dry chemical analysis system and the imported VITROS 4600 dry chemical analysis system. CONCLUSIONS: The verified precision, trueness, linear range, and inter-instrument comparison results for the four analytes fully met the established quality objectives, confirming that the QuidelOrtho VITROS 450 automatic biochemical analyzer is clinically applicable for the detection of these four analytes, thereby ensuring reliable results for routine laboratory testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Precision and Trueness Verification of Routine Biochemistry Tests using CLSI EP15-A3.Clinical laboratory · 2026
- Performance Evaluation of Dry Chemical Colorimetry for Plasma Ammonia Detection on the VITROS 350 Chemistry System.Journal of clinical laboratory analysis · 2026
- Multicenter evaluation of analytical performance and onboard stability of Mindray 10 concentrated biochemical reagentsJournal of Laboratory Medicine · 2026
- Analytical performance of a single epitope meta-sandwich total 25-hydroxyvitamin D immunoassay and establishment of reference intervals in a Chinese population.Annals of clinical biochemistry · 2026
- Analytical Performance Specifications Derived from External Quality Assessment Data: Comparison with Biological Variation-Based Models for Intermediate Precision Assessment.Annals of laboratory medicine · 2026
- B-302 Body Fluids as Complex Analytical Metrices: Challenges in Validation and Analytical Performance of Automated Biochemistry AnalyzersClinical Chemistry · 2026