Smartwatch-based loss-of-pulse detection for out-of-hospital cardiac arrest: the BECA-S2 and BECA-S6 studies
R. G. Hup, M. E. Kortman, R. W. F. H. Kuijpers, M. J. S. Mensinck, C. Bouchnaf, A. J. Ouss, H. L. Tan, R. Haakma,
Key points
From the paper's conclusion- These results show the feasibility of smartwatch-based loss-of-pulse detection.
Abstract
Objective. Out-of-hospital cardiac arrest (OHCA) is often unwitnessed, which delays treatment and lowers the chance of survival. A smartwatch could act as a witness by detecting loss of pulse from wrist photoplethysmography (PPG) and subsequently alert emergency services. We aimed to develop and evaluate such a detection model. Approach. As smartwatch data of OHCA are extremely difficult to acquire, we used in-hospital proxy data to test sensitivity: wrist PPG and accelerometry recorded during baseline and induced loss-of-pulse periods in 31 patients during catheter ablation of ventricular arrhythmias and defibrillation threshold (DFT) testing during subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation (BECA-S2). In total, 33 loss-of-pulse events, annotated using accompanying electrocardiograms and, where available, arterial blood pressure, were included. To count false alarms, we used wrist PPG and accelerometry of 100 healthy volunteers during unsupervised daily activities (BECA-S6). A recurrent neural network was evaluated with cross-validation, with one fold per patient with events. As the electrical VF induction and defibrillation shocks applied during these procedures would not be present in true OHCA episodes, we tested for and mitigated the resulting target leakage. Main results. The Area Under the Receiver Operating Characteristic Curve (AUC-ROC) was 86.8% (95% CI: 79.7 to 92.5%) and the Area Under the Precision-Recall Curve (PR-AUC) was 52.4% (95% CI: 38.6 to 66.2%). At a false alarm rate of zero on BECA-S6, the event-level sensitivity on BECA-S2 was 48.5% (95% CI: 32.1 to 65.5%). Higher sensitivities came at the cost of more false alarms. Without mitigation, target leakage inflated the model scores by 35.3% (95% CI: 11.5 to 61.4%) of the model's discriminative range. Significance. These results show the feasibility of smartwatch-based loss-of-pulse detection. Larger studies with data more representative of OHCA are needed. We recommend that future studies on induced loss of pulse test for and mitigate target leakage.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Wristband-based automated cardiac arrest detection and alerting: Algorithm performance in relation to the left ventricular functionEuropean Heart Journal · 2026
- Wearable devices for detection of cardiac arrest: a prospective proof-of-concept studyEuropean Heart Journal · 2026
- Noninvasive Cardiac Arrest Recognition Using Wearable Sensors in a Porcine Model of Ventricular Fibrillation.bioRxiv : the preprint server for biology · 2026
- Wearable-based automated cardiac arrest detection and alerting: A systematic review.Resuscitation · 2026
- Wearable saving lives: a case report.Resuscitation plus · 2026
- Validation of Commercial Wearables for Driver Monitoring: Assessing Heart Rate Accuracy and Reliability for Early Detection of Sudden IllnessesApplied Sciences · 2026