bioRxivPreprintOpen accessbiophysics

Hour-long live-cell TIRF imaging of GFP-talin at ensemble and single-molecule level with low-duty-cycle picosecond supercontinuum excitation

F. M. Margadant, O. Bakke, M. Sheetz, X. Hu

Key points

From the paper's conclusion
  • We describe an objective-type TIRF microscope illuminated by a picosecond supercontinuum laser spectrally windowed to 20-nm bands, pulse-picked to 1 MHz and delivered through two TIRF illuminators, one per polarisation direction.

Abstract

Photobleaching and phototoxicity limit how long living cells can be followed by total internal reflection fluorescence (TIRF) microscopy. We describe an objective-type TIRF microscope illuminated by a picosecond supercontinuum laser spectrally windowed to 20-nm bands, pulse-picked to 1 MHz and delivered through two TIRF illuminators, one per polarisation direction. Single GFP-talin molecules in focal adhesions of living cells were detected throughout uninterrupted recordings of 10-31 min, and drift-corrected, time-stamped localisations resolved adhesion motion of tens of nanometres per minute. In diffraction-limited time-lapse at one frame per second, adhesion brightness, adhesion area, cell footprint and adhesion motility were unchanged after an hour. Under continuous-wave excitation on a comparable commercial TIRF microscope, the same schedule preserved cell behaviour at a low laser setting but adhesions dimmed, and higher settings bleached and retracted cells within a minute of exposure. We provide the datasets and open ImageJ tools for drift correction and localisation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.