Towards an understanding of injury in aortic tissue prior to actual failure: quantification of mechanically induced collagen damage using collagen hybridizing peptides
A. Pukaluk, D. C. Haspinger, H. Wolinski, C. Viertler, J. C. Nawroth, G. A. Holzapfel
Key points
From the paper's conclusion- Therefore, we optimized an experimental workflow for large arteries that accounts for their energy-storage function and geometrical constraints; furthermore, we have characterized the tissue's autofluorescence and selected a suitable CHP fluorophore.
Abstract
The development, validation, and optimization of endovascular aortic devices require quantitative assessment of tissue damage occurring during surgical interventions. Until now, this assessment--before complete failure--has been largely unfeasible for lack of a suitable methodology. Recently developed collagen hybridizing peptides (CHPs) allow selective detection of damage to collagen, the primary load-bearing constituent of the aortic wall. However, their application in large, mechanically loaded arteries has not yet been established. Therefore, we optimized an experimental workflow for large arteries that accounts for their energy-storage function and geometrical constraints; furthermore, we have characterized the tissue's autofluorescence and selected a suitable CHP fluorophore. Specimens from a single porcine aorta were examined using uniaxial tensile testing, histological analysis, fluorescence spectral analysis, confocal microscopy, and quantitative image analysis to present collagen damage as a function of tissue stretch. The workflow detected load-induced collagen damage before tissue failure, whereas histological evaluation showed no obvious load-induced disruption. Damage scores were comparable between the undeformed control and the specimen stretched by a factor of 1.2, while specimens stretched beyond 1.2 exhibited elevated scores. This pattern corresponded to higher complication rates with excessive stent oversizing. Furthermore, the proposed protocol can be adapted for ex vivo stent deployment experiments to compare collagen damage resulting from various aortic device designs and deployment conditions. Quantitative damage metrics could inform structurally motivated constitutive material models for computational simulations of endovascular interventions. Combined with histological and structural analyses, this workflow could also help link tissue composition--or organization--to individual susceptibility to damage.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Strain-Stabilization of Collagen in Heart Valve Tissues: A Biomechanical Account of Proteolysis.Journal of biomechanical engineering · 2026
- Pericardium tissue layer and calcification characteristics - polarization-sensitive optical coherence tomography as material analysis tool for aortic valve bioprosthesis fabricationEuropean Heart Journal · 2026
- Analysis of Energy Dissipation Ratio in Commercial Bovine Pericardial Patches Treated with Glutaraldehyde Solution.Journal of functional biomaterials · 2026
- Mechanics of Balloon and Stent Angioplasty in Congenital Heart Interventions.Pediatric cardiology · 2026
- A Hybrid Hierarchical Framework for Quantifying Mechanical Markers in Atherosclerotic Disease Progression: A New Approach for Diagnosis and Risk Assessment.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026
- Association between multiaxial mechanical properties and microstructure of porcine descending aorta.Journal of the mechanical behavior of biomedical materials · 2026