Evaluation of sources of variability in a TEER (trans-epithelial/endothelial electrical resistance) system
R. Gutierrez, S. Lund, V. Reipa, E. J. Petersen
Key points
From the paper's conclusion- The findings from this study can influence how statistical analyses are performed on TEER results and can inform the design of robust TEER methods.
Abstract
Trans-epithelial/endothelial electrical resistance (TEER) measurements are a key nondestructive measurement of the barrier integrity of cells grown on inserts. To support the use of TEER measurements in bioassays, we have evaluated the sources of variability in a case study involving a chamber-based TEER system to assess a gut co-culture model. Data from 23 separate experiments, each lasting 22 d, were evaluated. Multiple lines of evidence suggested minimal variability within a run, a period during which several plates were analyzed sequentially. In contrast, run-to-run variability was substantially larger, even when two runs were performed several hours apart on the same day. Bayesian analysis of both KCl and blank insert data revealed that run-to-run variability was among the largest sources of variability, with relative uncertainty values of 3 % to 6 % of the mean value. Correlational analysis investigated whether deviations of KCl or blank insert TEER values within a run compared to long-term median values were correlated with deviations of the biculture gut model values within that same run from their long-term median values. Unexpectedly, the KCl values were not correlated, while the blank insert values had a weak correlation. We also evaluated the information gained from different control measurements. The findings from this study can influence how statistical analyses are performed on TEER results and can inform the design of robust TEER methods. For example, the approaches described here can be used to characterize variability within and among runs in TEER measurements for other methods.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Sensors Manufacturing on a 3D-Printed Transwell-Based Hybrid Organ-on-a-Chip for Non-Invasive Real-Time Biological Barrier Resistance MonitoringACS Biomaterials Science & Engineering · 2025
- Active Shielding of Glass Capillaries for the Compensation of Parasitic Effects in Transepithelial Electrical Resistance Measurements of OrganoidsCurrent Directions in Biomedical Engineering · 2026
- Methodological Variations in Cell Viability Assays: A Systematic Review and Experimental Evaluation Using a Caco-2 Cell ModelACS ES&T Toxicology · 2026
- Barrier function of the Vero E6 cell monolayer.Tissue barriers · 2026
- Assessment of epithelial barrier integrity in cellular monolayer models using low cost, clinically approved fluorescent contrast agents.PloS one · 2026
- Electrochemical impedance spectroscopy supporting live imaging for barrier assessment in a microphysiological system.Bioelectrochemistry (Amsterdam, Netherlands) · 2026