Differential sensing with imprinted and non-imprinted organic electrochemical transistors reduces calibration drift in wearable sweat cortisol sensors
L. I. Almofeez, M. S. Islam, T. S. Ambagahawaththa, Y. Khan
Key points
From the paper's conclusion- In sweat from one participant, read against the calibration from day 1, both outputs tracked the direction of visible changes in salivary cortisol, while the MIP alone diverged from the differential output, consistent with the laboratory drift.
Abstract
The output of a wearable cortisol sensor drifts over days for reasons unrelated to cortisol, so a calibration made on the first day soon stops being valid. In our differential organic electrochemical transistor, a PEDOT:PSS transistor coated with a molecularly imprinted polymer (MIP) for cortisol is paired with one coated with a non-imprinted polymer (NIP) of the same composition, and the NIP response is subtracted. Over four calibrations in artificial sweat across 14 days, the response of the MIP alone shifted by 15% to 22% of its range on day 1, whereas the differential output shifted by 1% to 10%, and the variation between days at 1 to 10nM fell by 65% to 70% on average for individual pairs. In sweat from one participant, read against the calibration from day 1, both outputs tracked the direction of visible changes in salivary cortisol, while the MIP alone diverged from the differential output, consistent with the laboratory drift.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Wearable Organic Electrochemical Transistor Biosensor for Real-Time Measurement of Sweat Salinity and Saturation.ACS omega · 2026
- Redox-Active Molecularly Imprinted Polymer: Synergistic Integration of Recognition and Signal Transduction for Sweat Cortisol SensingAnalytical Chemistry · 2026
- Self-Correction of pH-Induced Signal Variations in Methylene Blue-Labeled Aptamer Electrochemical Biosensors: Wearable Cortisol Detection in SweatACS Sensors · 2026
- A multimodal wearable microfluidic platform for in situ sweat cortisol monitoring with a NiHCF-enabled molecularly imprinted electrochemical sensor.Biosensors & bioelectronics · 2026
- Wearable Microfluidic Chip Integrated with Chrono-Sampling for Cortisol Detection in Sweat.Analytical chemistry · 2026
- An Electrochemical Cortisol Sensor Based on rGO-Modified Molecularly Imprinted PolymersJournal of Composites Science · 2026