Validation of an electronic-nose sensor array in detecting C. difficile and differentiating toxin-positive from toxin-negative stool samples
P. N. Balius, C. Campbell, M. Ikrma, H. H. Nguyen
Key points
From the paper's conclusion- Conclusion: An e-nose sensor array partially differentiated PCR-classified C. difficile specimen classes at the exposure level under headspace-controlled conditions.
Abstract
Objective: To evaluate the specimen-level performance of a multichannel electronic nose (e-nose) sensor array in classifying stool specimens into relevant Clostridioides difficile categories. Design: Retrospective single-center validation study with reference to polymerase chain reaction (PCR) and C. difficile toxin testing. Participants: A convenience sample of 31 discarded stool specimens with PCR and confirmatory toxin testing. Categorized into PCR-positive (+)/toxin+, PCR+/toxin-negative (-), and PCR-. Methods: Specimens were sampled in controlled headspace by two e-nose sensor arrays in 10-minute sessions with 5-second sampling intervals. The primary endpoint was one-versus rest classification performance (sensitivity, specificity, positive predictive value, and accuracy) into three classes: PCR+/toxin+, PCR+/toxin-, and PCR-. Data were normalized and the region of interest was extracted for training and testing. Random forests were utilized for classification of normalized data with a test-train split at the specimen-level. Results: There were 457 unique 10-minute sampling events (147 PCR+/toxin+, 183 PCR+/toxin-, 127 PCR-). Sensors 1 and 2 captured 20,331 and 23,538 unique 5-second exposures within the region of interest, respectively. Testing data classification accuracy was 0.79 [0.78, 0.81] and 0.84 [0.83, 0.85] for Sensors 1 and 2. Sensitivity for PCR+/toxin+ exposures was 0.82 [0.79, 0.84] and 0.94 [0.92, 0.95], and specificity was 0.91[0.90, 0.93] and 0.93 [0.92, 0.94] for Sensors 1 and 2. Conclusion: An e-nose sensor array partially differentiated PCR-classified C. difficile specimen classes at the exposure level under headspace-controlled conditions. Performance was strongest for PCR+/toxin+ exposures. Findings support further prospective validation in the hospital room setting.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Integration of Electronic Nose Profiles from Breath, Urine, and Axillary Sweat for Colorectal Cancer ClassificationChemosensors · 2026
- Machine Learning-Assisted Experimental Electronic Nose Measurements for Biopesticide Treatment Discrimination in Tomatoes.Sensors (Basel, Switzerland) · 2026
- Rapid Identification of Foodborne Pathogen Biofilms Using Fluorescence Sensor Array and Machine Learning ModelsACS Food Science & Technology · 2026
- A High-Performance Artificial Olfactory Chip for Real-Time Cold Chain Food Freshness MonitoringACS Nano · 2026
- Relationship Between Clostridioides difficile Stool Concentration, Nucleic Acid Amplification Test Results, and Environmental ContaminationOpen Forum Infectious Diseases · 2026
- Clinical characteristics of ICU patients with gastrointestinal colonisation by Candida auris.Frontiers in cellular and infection microbiology · 2026