Key points

From the paper's conclusion
  • In patients with suspected IVA, long-term Smartpatch monitoring demonstrated higher arrhythmia detection rates, improved diagnostic accuracy, enhanced HRV assessment, and better patient comfort compared with 24-hour Holter monitoring. These findings support its use as a valuable tool for screening and diagnosis in patients with suspected IVA.

Abstract

OBJECTIVE: To compare the diagnostic performance, patient comfort, and heart rate variability (HRV) assessment of a novel long-term Smartpatch monitor with those of a conventional 24-hour Holter monitor in patients with suspected idiopathic ventricular arrhythmia (IVA). METHODS: This retrospective self-controlled study enrolled 164 patients who underwent simultaneous 24-hour Holter and 7- to 14-day Smartpatch monitoring. Detection rates of arrhythmia subtypes, premature ventricular contraction (PVC) burden, symptom-rhythm correlation, wearing comfort (assessed using 5-point Likert scale), HRV indices (SDNN, SDANN, RMSSD), and diagnostic accuracy were compared between different monitoring methods. RESULTS: Compared with Holter monitoring, Smartpatch showed significantly higher detection rates for all arrhythmia subtypes (all P < 0.05). Among the 85 patients with negative Holter findings, Smartpatch provided an incremental diagnostic yield of 18.8% (16/85), resulting in changes in clinical management for 9 patients (5 receiving pharmacological therapy and 4 undergoing ablation). Smartpatch was associated with superior overall comfort scores (4.3 ± 0.5 vs 2.9 ± 0.9, P = 0.007), a lower premature termination rate (1.2% vs 5.5%, P = 0.035), a higher proportion of valid recording data (98.5% vs 96.2%, P < 0.001), and a mean wear duration of 13.2 days. HRV indices were significantly higher with Smartpatch monitoring than with Holter monitoring (P < 0.05). Diagnostic sensitivity, specificity, and accuracy for IVA were 94.2%, 91.8%, and 93.1% with Smartpatch, respectively, compared with 78.5%, 85.4%, and 80.2% with Holter (P < 0.001, P = 0.013, P < 0.001, respectively). CONCLUSION: In patients with suspected IVA, long-term Smartpatch monitoring demonstrated higher arrhythmia detection rates, improved diagnostic accuracy, enhanced HRV assessment, and better patient comfort compared with 24-hour Holter monitoring. These findings support its use as a valuable tool for screening and diagnosis in patients with suspected IVA.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Smartpatch long-term monitoring superior to 24-hour Holter for detecting idiopathic ventricular arrhythmia: a retrospective self-controlled study — Science Skim