Smartpatch long-term monitoring superior to 24-hour Holter for detecting idiopathic ventricular arrhythmia: a retrospective self-controlled study.
Lutao Lin, Xiaodong Zheng, Zhuodong Zhou, Yixuan Liu, Yongfu Zhao
Department of Cardiology, The First People's Hospital of Hechi City Hechi 546300, Guangxi, China. · Department of Cardiology, Guangxi Xing'an County People's Hospital Guilin 541300, Guangxi, China.
Key points
From the paper's conclusion- In patients with suspected IVA, long-term Smartpatch monitoring demonstrated higher arrhythmia detection rates, improved diagnostic accuracy, enhanced HRV assessment, and better patient comfort compared with 24-hour Holter monitoring. These findings support its use as a valuable tool for screening and diagnosis in patients with suspected IVA.
Abstract
OBJECTIVE: To compare the diagnostic performance, patient comfort, and heart rate variability (HRV) assessment of a novel long-term Smartpatch monitor with those of a conventional 24-hour Holter monitor in patients with suspected idiopathic ventricular arrhythmia (IVA). METHODS: This retrospective self-controlled study enrolled 164 patients who underwent simultaneous 24-hour Holter and 7- to 14-day Smartpatch monitoring. Detection rates of arrhythmia subtypes, premature ventricular contraction (PVC) burden, symptom-rhythm correlation, wearing comfort (assessed using 5-point Likert scale), HRV indices (SDNN, SDANN, RMSSD), and diagnostic accuracy were compared between different monitoring methods. RESULTS: Compared with Holter monitoring, Smartpatch showed significantly higher detection rates for all arrhythmia subtypes (all P < 0.05). Among the 85 patients with negative Holter findings, Smartpatch provided an incremental diagnostic yield of 18.8% (16/85), resulting in changes in clinical management for 9 patients (5 receiving pharmacological therapy and 4 undergoing ablation). Smartpatch was associated with superior overall comfort scores (4.3 ± 0.5 vs 2.9 ± 0.9, P = 0.007), a lower premature termination rate (1.2% vs 5.5%, P = 0.035), a higher proportion of valid recording data (98.5% vs 96.2%, P < 0.001), and a mean wear duration of 13.2 days. HRV indices were significantly higher with Smartpatch monitoring than with Holter monitoring (P < 0.05). Diagnostic sensitivity, specificity, and accuracy for IVA were 94.2%, 91.8%, and 93.1% with Smartpatch, respectively, compared with 78.5%, 85.4%, and 80.2% with Holter (P < 0.001, P = 0.013, P < 0.001, respectively). CONCLUSION: In patients with suspected IVA, long-term Smartpatch monitoring demonstrated higher arrhythmia detection rates, improved diagnostic accuracy, enhanced HRV assessment, and better patient comfort compared with 24-hour Holter monitoring. These findings support its use as a valuable tool for screening and diagnosis in patients with suspected IVA.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Detection of arrhythmias by 24-hour holter monitoring versus 14-day continuous electrocardiogram patch: a systematic review and meta-analysisEuropean Heart Journal · 2026
- Automatic measurement of short-term variability of repolarization on human intracardiac electrograms predicts VT/VF events.Heart rhythm O2 · 2026
- Patient experiences with a smartwatch 1L-ECG versus traditional Holter monitoring for ambulatory cardiac rhythm monitoring: a qualitative studyBMJ Open · 2025
- Real-world evaluation of the positive predictive value and transmission burden of a novel implantable cardiac monitor with a remote reprogramming capability: insights from a single-center experienceFrontiers in Cardiovascular Medicine · 2026
- Comparing the Feasibility of S-Patch vs 24-Hour Holter in Frail Older People With Falls: Randomized Open-Label Crossover Pilot Study.JMIR rehabilitation and assistive technologies · 2026
- Arrhythmic mitral valve prolapse detection using long-term ambulatory rhythm monitoring - the ALARM-PILOT studyEuropean Heart Journal · 2026