Post-Detachment Displacement of a Woven EndoBridge (WEB) Device Managed via Microcatheter Manipulation: A Case Report.
Kyoichi Seo, Yoji Kuramoto, Soichiro Abe, Shinichi Yoshimura
Department of Neurosurgery, Hyogo Medical University, Nishinomiya, Hyogo, Japan.
Key points
From the paper's conclusion- Post-detachment WEB displacement can occur due to dynamic vascular shifts even with precise volumetric sizing. Microcatheter manipulation is a feasible alternative bailout option that preserves the low-profile advantages of the WEB device and avoids the risks associated with rescue stenting.
Abstract
OBJECTIVE: To report a case of post-detachment Woven EndoBridge (WEB) displacement successfully managed through microcatheter manipulation alone, avoiding unplanned stenting, which necessitates long-term dual antiplatelet therapy. CASE PRESENTATION: A 42-year-old male presented with an unruptured middle cerebral artery bifurcation aneurysm. Device sizing was planned using high-resolution 3D DSA and volumetric analysis on a Siemens ARTIS icono biplane (ARTIS icono VE20 software; Siemens Healthineers, Erlangen, Germany). A WEB SL 5 × 3 mm device (Terumo Neuro, Tustin, CA, USA) was deployed and detached. However, the subsequent withdrawal of an adjunctive microcatheter from a branch vessel caused vascular recoil, leading the detached WEB to rotate and protrude toward the branch orifice, compromising distal flow. Instead of rescue stenting, we performed a "pushing technique" using a Leonis Mova microcatheter (SB-Kawasumi Laboratories, Kanagawa, Japan). The device was successfully reseated within the aneurysm sac, albeit with a 90° rotation. Final angiography confirmed stable seating and excellent branch patency. The patient was discharged without neurological deficits on single antiplatelet therapy. CONCLUSION: Post-detachment WEB displacement can occur due to dynamic vascular shifts even with precise volumetric sizing. Microcatheter manipulation is a feasible alternative bailout option that preserves the low-profile advantages of the WEB device and avoids the risks associated with rescue stenting.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Rare delayed thromboembolic complication 18 days after Woven EndoBridge implantation: A case report.Surgical neurology international · 2026
- Repeat Woven EndoBridge Implantation for Recurrent Intracranial Aneurysms: Experience from a Consecutive Single-Center Cohort.Clinical neuroradiology · 2026
- Successful Percutaneous Retrieval of an Embolized MitraClip From the Right Subclavian Artery.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Impact of Woven EndoBridge (WEB) device protrusion on postprocedural diffusion-weighted imaging lesions.Journal of neurointerventional surgery · 2026
- Recovery of oculomotor nerve palsy following Woven EndoBridge treatment for a symptomatic basilar artery-superior cerebellar artery junction aneurysm: illustrative case.Journal of neurosurgery. Case lessons · 2026
- 9V battery-powered microguidewire electrocoagulation and flow diversion for salvage of recurrent supraclinoid aneurysm.Journal of neurointerventional surgery · 2026