Diagnostic Performance of Combined Supine-Prone Quantitative Myocardial Perfusion SPECT for Detecting Obstructive Coronary Artery Disease: A Patient- and Vessel-Territory-Based Retrospective Study.
Mehmet Ateş, Furkan Gür, Faruk Recep Özalp, Emre Paçacı, Fadime Topalhan
Department of Nuclear Medicine, Osmaniye Training and Research Hospital, Osmaniye, Türkiye. · Department of Medical Oncology, Osmaniye Training and Research Hospital, Osmaniye, Türkiye.
Key points
From the paper's conclusion- In male patients undergoing a complete stress-rest examination, software-derived combined supine-prone quantitative myocardial perfusion SPECT (MPS) improved specificity and overall diagnostic accuracy compared with supine-only assessment by reducing false-positive findings caused by inferior and inferolateral attenuation. Combined positioning should be viewed as an adjunct to, rather than a substitute for, paired stress-rest assessment and physician visual interpretation.
Abstract
OBJECTIVE: To assess and compare the diagnostic performance of supine-only, prone-only, and software-derived combined supine-prone quantitative myocardial perfusion single-photon emission computed tomography (SPECT) for identifying obstructive coronary artery disease in male patients. METHOD: This retrospective, single-center diagnostic accuracy study evaluated men who underwent a one-day stress-rest myocardial perfusion SPECT protocol with technetium-99m (Tc-99m) sestamibi, had both supine and prone stress and rest acquisitions, and subsequently underwent invasive coronary angiography within 3 months. Total perfusion deficit (TPD), summed stress score (SSS), summed rest score (SRS), summed difference score (SDS), and vessel-territory measurements were recorded from the supine, prone, and combined datasets. Obstructive coronary artery disease was defined as stenosis of at least 50% in the left main coronary artery or at least 70% in the left anterior descending artery (LAD), left circumflex artery (LCx), or right coronary artery (RCA). RESULTS: Combined stress TPD demonstrated the greatest patient-level discriminatory ability (area under the curve [AUC], 0.892; 95% confidence interval [CI], 0.846-0.938). The corresponding AUCs for supine and prone stress TPD were 0.798 and 0.868, respectively. The AUC gain for combined versus supine stress TPD was 0.094 (95% CI, 0.048-0.140; P < 0.001), whereas the gain for combined versus prone stress TPD was 0.024 (95% CI, -0.008 to 0.056; P = 0.142). For supine stress TPD, the optimal cutoff was 6.20, yielding 77.0% sensitivity, 73.7% specificity, and 74.5% accuracy. For combined stress TPD, the optimal cutoff was 4.10, yielding 88.5% sensitivity, 84.4% specificity, and 85.4% accuracy. CONCLUSION: In male patients undergoing a complete stress-rest examination, software-derived combined supine-prone quantitative myocardial perfusion SPECT (MPS) improved specificity and overall diagnostic accuracy compared with supine-only assessment by reducing false-positive findings caused by inferior and inferolateral attenuation. Combined positioning should be viewed as an adjunct to, rather than a substitute for, paired stress-rest assessment and physician visual interpretation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Platform-specific impact of CT attenuation correction on CZT myocardial perfusion SPECT: per-vessel detection of obstructive coronary artery disease.The international journal of cardiovascular imaging · 2026
- Discordance Between Gated Single-Photon Emission Computed Tomography Myocardial Perfusion Imaging and Coronary Angiography in Dilated Cardiomyopathy.Cureus · 2026
- D-SPECT combined with deep learning predicts obstructive coronary artery disease.Frontiers in cardiovascular medicine · 2026
- Diffusely impaired myocardial perfusion in patients with suspected coronary artery disease: Clinical utility of relative perfusionClinical Physiology and Functional Imaging · 2026
- Comparative analysis of photon-counting and energy-integrating detector CT to identify obstructive coronary artery diseaseEuropean Radiology · 2025
- Assessment of myocardial perfusion defects with first-pass arterial-phase spectral photon counting CTEuropean Heart Journal · 2026