Atmospheric Measurement Techniques

A Compact Fabry-Perot Interferometer with a Solid Etalon for Measuring Upper Atmospheric Wind and Temperature

Tiancai Wang, Yajun Zhu, Guangyi Zhu, Wei Yuan, Zhongqi Tan, Weijun Liu, Qiucheng Gong, Hong Gao,

Key points

From the paper's conclusion
  • A compact solid-etalon Fabry-Perot interferometer (SEFPI) with a 70 mm-diameter solid fused-silica etalon has been developed for ground-based measurements of upper atmospheric winds and temperatures at an altitude of approximately 250 km.

Abstract

A compact solid-etalon Fabry-Perot interferometer (SEFPI) with a 70 mm-diameter solid fused-silica etalon has been developed for ground-based measurements of upper atmospheric winds and temperatures at an altitude of approximately 250 km. Unlike conventional FPIs employing air-spaced etalons, the SEFPI uses a solid etalon, enabling a more compact, lightweight, and cost-effective instrument design, while requiring stringent thermal stabilization. Laboratory validation using an acousto-optic frequency shifter (AOFS) to generate controlled Doppler shifts demonstrated the SEFPI's high measurement accuracy, with an average wind uncertainty of 0.42 m s −1 and a root mean square error (RMSE) of 0.61 m s −1 . Preliminary field validation and intercomparisons were conducted via coincident ground-based measurements with two interferometry techniques. The first employed a Dual-Channel Optical Interferometer (DCOI), while the second used a well-established FPI employing a 100 mm-diameter air-spaced etalon. The SEFPI measurements exhibited strong agreement with both instruments, yielding Pearson correlation coefficients exceeding 0.84. Statistical analysis yielded a mean difference (DCOI/FPI – SEFPI), averaged across both meridional and zonal wind components, of +1.88 m s −1 relative to the co-located DCOI instrument and +5.58 m s −1 relative to the conventional FPI located 502 km away. Temperature comparisons between the two FPIs revealed a correlation coefficient of 0.86, with a mean difference (FPI – SEFPI) of +62.38 K.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers