EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology

Safety and effectiveness of the Vertex Pulmonary Embolectomy System in acute intermediate-risk pulmonary embolism: the SPIRARE II pivotal trial.

Sameh Sayfo, Jakub Stępniewski, Vincent Gallo, Joseph S Rossi, Irene Marthe Lang, Joseph Ingrassia, Russell D Rosenberg, Rohit Amin,

Baylor Scott & White Heart Hospital, Plano, TX, USA. · Pulmonary Circulation Centre, Department of Cardiac and Vascular Diseases, Jagiellonian University Medical College, John Paul II Hospital, Krakow, Poland.

Trial · n=123

Key points

From the paper's conclusion
  • The Vertex Pulmonary Embolectomy System met its prespecified safety and effectiveness endpoints in acute intermediate-risk PE. TFX delivered precise, stable control with reliable vessel access, enabling consistent right ventricular unloading. (ClinicalTrials.gov: NCT06576427).

Abstract

BACKGROUND: Percutaneous aspiration thrombectomy can rapidly reduce afterload in acute pulmonary embolism (PE), but catheter stiffness and instability may limit access and prevent effective right ventricle (RV) unloading. The Vertex Pulmonary Embolectomy System with Transforming Fixation (TFX) technology is designed to address current limitations of existing catheter-based systems by separating the flexibility necessary for safe navigation from the support that is essential for procedural control. AIMS: We aimed to evaluate the safety and effectiveness of the Vertex Pulmonary Embolectomy System for mechanical aspiration thrombectomy in acute intermediate-risk PE. METHODS: SPIRARE II was a prospective, single-arm, multicentre pivotal investigation conducted at 19 sites in the United States and Europe. Eligible patients had acute PE (≤14 days), a computed tomography angiography-assessed RV to left ventricle (RV/LV) ratio>0.9, and haemodynamic stability (systolic blood pressure ≥90 mmHg). Co-primary endpoints were composite major adverse events (MAE) within 48 hours and core laboratory-assessed RV/LV change from baseline to 48 hours. Prespecified secondary endpoints were collected immediately pre- and post-procedure in all patients including mean pulmonary artery pressure, total pulmonary vascular resistance, and cardiac index. RESULTS: In 123 patients, the 48-hour composite MAE rate was 2.4% (3/123), meeting the prespecified performance goal (one-sided p<0.0001). The mean RV/LV diameter ratio decreased from 1.44±0.33 at baseline to 1.05±0.21 at 48 hours, a mean reduction of 0.39±0.33, exceeding the prespecified performance goal (p<0.0001). CONCLUSIONS: The Vertex Pulmonary Embolectomy System met its prespecified safety and effectiveness endpoints in acute intermediate-risk PE. TFX delivered precise, stable control with reliable vessel access, enabling consistent right ventricular unloading. (ClinicalTrials.gov: NCT06576427).

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Safety and effectiveness of the Vertex Pulmonary Embolectomy System in acute intermediate-risk pulmonary embolism: the SPIRARE II pivotal trial — Science Skim