Technical Performance and Automated-manual Scoring Agreement in Type III Home Sleep Apnea Testing: A Single-center Study from Saudi Arabia.
Sultan Aldhabei, Raghad Alhajaji, Salih A Aleissi, Hamza O Dhafar, Samar Z Nashwan, Mohammad Alrashed, Abdullah Aldhuwaihy, Talal Alghadir,
Department of Medicine, College of Medicine, Najran University, Najran, Saudi Arabia. · Sleep Medicine Department, King Saud University Medical City, Riyadh, Saudi Arabia.
Key points
From the paper's conclusion- Among adults from Saudi Arabia, type III HSAT had a high technical success rate and strong overall agreement between automated and manual scoring. These findings support the use of HSAT as a reliable diagnostic alternative in appropriately selected adults with a high pretest probability of OSA.
Abstract
BACKGROUND: Obstructive sleep apnea (OSA) is a prevalent disorder that carries significant health risks. Home sleep apnea testing (HSAT) is a cost-effective alternative to in-lab polysomnography for diagnosing OSA. OBJECTIVES: To evaluate the technical performance and automated-manual scoring agreement of type III HSAT in a Saudi population with a high pretest probability of OSA, and to establish a local performance baseline that can inform future multicenter and cross-regional comparisons. METHODS: This retrospective study included adults with a high pretest probability of moderate-severe OSA who underwent type III HSAT at a tertiary sleep center in Riyadh, Saudi Arabia. Automated HSAT recordings (Nox T3 device) were re-scored manually by a blinded sleep technologist. Key outcomes included technical success and agreement between automated and manual scoring of respiratory indices. RESULTS: A total of 173 participants were included, of which 128 (74.4%) had OSA. HSAT was technically successful in 90.7% of cases. Automated apnea-hypopnea index demonstrated high overall concordance with manual scoring (CCC = 0.919; 95% CI: 0.893-0.940), with minimal mean bias. The apnea index (CCC = 0.993; 95% CI: 0.990-0.995) and oxygen desaturation index (CCC = 0.973; 95% CI: 0.964-0.980) demonstrated excellent concordance, while the hypopnea index showed moderate concordance (CCC = 0.820; 95% CI: 0.764-0.864) with modest automated underestimation. In technically failed cases, nasal airflow showed the poorest signal quality (mean ± SD, 58.4% ± 27.4%) followed by pulse oximetry (86.7% ± 25.1%). CONCLUSION: Among adults from Saudi Arabia, type III HSAT had a high technical success rate and strong overall agreement between automated and manual scoring. These findings support the use of HSAT as a reliable diagnostic alternative in appropriately selected adults with a high pretest probability of OSA.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Modified scoring criteria to improve the accuracy of the home sleep apnea testSleep And Breathing · 2026
- A Comparative Review of Performance and Concurrent Cardiac Rhythm Assessment Across Home Sleep Apnea Testing Devices.Cureus · 2026
- Diagnostic Accuracy of the Oxygen Desaturation Index for Screening Severe Obstructive Sleep ApneaPakistan Journal of Health Sciences · 2026
- Deep Learning-Based Automated Versus Manual Snoring Scoring in Sleep Disordered Breathing.The Laryngoscope · 2026
- Metrics Reflecting Cumulative Exposure to Nocturnal Hypoxia Demonstrate Promising Performance in Distinguishing between Moderate and Severe Obstructive Sleep Apnea.Sleep science (Sao Paulo, Brazil) · 2026
- P012. Agreement between automated and manual scoring in children ≥13 yearsSLEEP advances · 2026