Assessment of Clinical Outcomes and Electrophysiological Parameters of Left Bundle Branch Area Pacing: A Single-center Prospective Study with 12-month Follow-up.
Naved Aslam, Anshuman Gupta, Bhavuk Jaiswal, Gurpreet Singh Wander, Shibba Takkar Chhabra, Abhishek Goyal, Akash Batta, Bishav Mohan
Department of Cardiology, Dayanand Medical College and Hospital, Ludhiana, Punjab, India. · Department of Cardiology, Fortis Hospital, Faridabad, Haryana, India.
Key points
From the paper's conclusion- With improvements observed in electrophysiological parameters, cardiac function, and minimal adverse events over a 12-month follow-up period, LBBP demonstrates both efficacy and safety in this patient population. These results advocate for further exploration and utilization of LBBP as a preferred pacing strategy in clinical practice.
Abstract
BACKGROUND: Left bundle branch pacing (LBBP) is developing as an effective technique for cardiac pacing, offering potential advantages over traditional right ventricular pacing methods. However, comprehensive assessment of LBBP outcomes across various indications and over extended follow-up periods is limited. This single-center prospective study aims to address this gap by evaluating the electrophysiological, echocardiographic, and clinical outcomes of LBBP over a 12-month follow-up period. METHODS: This was a single-center prospective cohort study with repeated follow-up assessments performed at 3, 6, and 12 months following left bundle branch area pacing (LBBAP) implantation. This study enrolled 63 patients aged 18-80 with symptomatic bradyarrhythmia and left ventricular dysfunction undergoing LBBAP using Select Secure pacing leads. Outcomes were compared pre- and post-implantation at 3-, 6-, and 12-month follow-up, assessing electrophysiological parameters, cardiac function, and adverse events. RESULTS: In this study, successful implantation was achieved in all patients, with selective LBB capture demonstrated in 78.9%. Lead parameters such as capture threshold, paced QRS duration, pacing impedance, and R wave amplitude at 3, 6, and 12 months when compared with the baseline, remained stable with no significant changes over time. Individuals with lower baseline ejection fraction EF improved significantly from baseline between 3 and 6 months, 6 and 12 months, and 3 and 12 months, respectively. Procedural and late complications were minimal. CONCLUSION: With improvements observed in electrophysiological parameters, cardiac function, and minimal adverse events over a 12-month follow-up period, LBBP demonstrates both efficacy and safety in this patient population. These results advocate for further exploration and utilization of LBBP as a preferred pacing strategy in clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Efficacy and safety of left bundle branch area pacing performed with stylet-driven leads.Frontiers in cardiovascular medicine · 2026
- Left Bundle Branch Area Pacing: Current Clinical Practice and Future Perspectives.Hellenic journal of cardiology : HJC = Hellenike kardiologike epitheorese · 2026
- Left bundle branch area pacing in children: a systematic review and meta-analysis of feasibility, safety, and outcomes.Frontiers in cardiovascular medicine · 2026
- Step-By-Step Left Bundle Branch Area Pacing: A Practical Technical Guide.Turk Kardiyoloji Dernegi arsivi : Turk Kardiyoloji Derneginin yayin organidir · 2026
- Left bundle branch area pacing outcomes from an African registry: Short-term findings from the MACOP study.Heart rhythm O2 · 2026
- Comparative clinical outcomes of stylet-driven left bundle branch area pacing versus right ventricular pacing in patients with preserved ejection fraction: A 3-year follow-up study.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026