Bidirectional High-Irradiance Light Emitting Diode Phototherapy Using Two Devices for Severe Neonatal Hyperbilirubinemia.
Ryutaro Sugimura, Sayo Morimoto, Yukihito Imagawa, Yu Masuda, Yuki Nakata, Takumi Kido, Mariko Ashina, Kazumichi Fujioka
Department of Pediatrics, Kobe University Graduate School of Medicine. · Department of Nursing, Kobe University Hospital.
Key points
From the paper's conclusion- BHIP may represent an additional therapeutic option for severe neonatal jaundice that cannot be adequately controlled with conventional intensive phototherapy, as a temporary rescue strategy while exchange transfusion is being considered or prepared.
Abstract
Severe neonatal hyperbilirubinemia can lead to acute bilirubin encephalopathy and kernicterus and may require exchange transfusion when conventional phototherapy is insufficient. Although exchange transfusion is an effective treatment, it is invasive, requires substantial personnel and resources, and is associated with potentially serious complications. Increasing irradiance and the body surface area exposed to light may provide a means of intensifying phototherapy when conventional treatment is insufficient. Here, this article describes a protocol for bidirectional high-irradiance phototherapy (BHIP) using two commercially available LED phototherapy systems. The overhead LED phototherapy device is positioned above the infant, while the compact and flexibly positionable adjustable LED phototherapy device is placed obliquely from either side, allowing irradiation from two directions and increasing the body surface area exposed to therapeutic light. Both devices are positioned at a distance of 30 cm from the infant. Irradiance is measured at four sites on the infant's body surface, and device positioning is adjusted to achieve a mean irradiance of 80-100 µW/cm2/nm. During treatment, cardiorespiratory status, body temperature, hydration, skin condition, eye protection, and signs of acute bilirubin encephalopathy are closely monitored. BHIP should not delay exchange transfusion when signs of acute bilirubin encephalopathy are present. This protocol provides a practical method for increasing both irradiance and exposed body surface area using existing LED phototherapy devices. BHIP may represent an additional therapeutic option for severe neonatal jaundice that cannot be adequately controlled with conventional intensive phototherapy, as a temporary rescue strategy while exchange transfusion is being considered or prepared. BHIP should not delay exchange transfusion when ET is clinically indicated.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Maintenance of Dose-Dependent Bilirubin Reduction with Ultra-High-Irradiance Phototherapy up to 148 μW/cm2/nm in Severe Neonatal Jaundice.Children (Basel, Switzerland) · 2026
- Therapeutic efficacy of intermittent versus continuous phototherapy in term neonates with hyperbilirubinemia: A randomized controlled trial.Journal of neonatal-perinatal medicine · 2026
- Solar-powered phototherapy for severe neonatal hyperbilirubinemia in low-resource settings: a multicentre implementation study in Nigeria.Frontiers in pediatrics · 2026
- Brief intensive phototherapy for newborns – benefits, risks, and pending questionsPediatric Research · 2026
- Intermittent phototherapy in neonatal hyperbilirubinemia: A pilot randomized controlled trial in Taiwan.Journal of the Formosan Medical Association = Taiwan yi zhi · 2026
- Emergency management of extreme hyperbilirubinemia.Seminars in perinatology · 2026