Remote Intraoperative Monitoring of Cochlear Implants.
Pascal Nachtigäller, Joachim Müller, Daniel Polterauer-Neuling
Division of Audiology, Department for Otorhinolaryngology, LMU University Hospital Munich. · Department for Otorhinolaryngology, LMU University Hospital Munich.
Key points
From the paper's conclusion- This protocol describes how cochlear implant measurements can be performed remotely using a telephone in the operating room and a remotely accessible laptop.
Abstract
Intraoperative measurements are valuable tools in cochlear implant (CI) surgery, enabling the determination of implant functionality while the patient is still in the operating room. Standard measurements include impedances for stimulating and non-stimulating contacts, triggering of the stapedial reflex with or without estimating the threshold, and assessing cochlear nerve function by measuring electrically evoked compound action potentials. Additional measurements can also be performed. For example, electrocochleography can be used to observe and help preserve residual hearing in patients. Usually, measurements are performed in situ in the operating room, which is time-consuming, especially for the engineer responsible for the measurements. This protocol describes how cochlear implant measurements can be performed remotely using a telephone in the operating room and a remotely accessible laptop. With this setup, the measurements can be completed without the engineer entering the operating room, reducing the time required from around 29 min in situ. to 8 min remotely (excluding session with ECochG).
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Remote CI Fitting in Early Rehabilitation Phase: Preliminary ResultsOtology & Neurotology · 2025
- Remote Check in Cochlear Implant Follow-Up: An Exploratory Comparison with Conventional Audiological Measurements.Audiology research · 2026
- Noninvasive indirect intracochlear pressure monitoring during cochlear implantationHearing Research · 2026
- Impedance-based tracking of a cochlear implant electrode reliably detects successful array insertion.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026
- Guiding Cochlear Implant Electrode Insertion Using Impedance Measurements: A Computational Simulation Study.Sensors (Basel, Switzerland) · 2026
- Intraoperative non-contact three-dimensional measurement of the facial recess using a 4K3D exoscope during cochlear implantation.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026