Agreement between spirometric quality reviewers in an epidemiological study.
Santiago C Arce, Ricardo Del Olmo, Adrián Ceccato, Juan M Luchelli, Evelyn Sureda, Mariano Fielli, Sergio J Arias, Andrés L Echazarreta,
. Instituto de Investigaciones Médicas Alfredo Lanari, Universidad de Buenos Aires, Ciudad Autónoma de Buenos Aires, Argentina. · . Hospital María Ferrer, Ciudad Autónoma de Buenos Aires, Argentina.
Key points
From the paper's conclusion- Reviewer variability significantly affects spirometric acceptability, with numerical criteria achieving higher agreement than subjective assessments. These discrepancies have implications for large-scale studies, data quality, and automated algorithm development. Cross-checking reviews is recommended to enhance reliability. Further standardization of acceptability criteria is warranted.
Abstract
OBJECTIVE: . Spirometry is the gold standard among pulmonary function tests, with acceptability criteria established to ensure reliable results. However, variability in the interpretation of spirometry by trained reviewers remains unexplored. This study assessed agreement among trained reviewers regarding spirometric acceptability and its components, including the number of acceptable maneuvers, predominant defect, and repeatability. METHODS: . Spirometries from 3,982 adults (≥40 years) participating in an epidemiological study across six cities in Argentina were independently reviewed by junior reviewers (JRs) and adjudicated by senior reviewers (SRs), following the 2005 ATS/ERS guidelines. Agreement was analyzed using Cohen's kappa, McNemar's test, and proportion comparisons. RESULTS: . A total of 7,964 pre- and post-bronchodilator spirometries were analyzed. Agreement on overall acceptability between JRs was 76% and 84% in two independent reviewer teams (RvTa and RvTb, respectively). Agreement on specific defects was highest for numerical criteria, such as insufficient expiratory time (k=0.61) and excessive back-extrapolated volume (BEV; k=0.52), but lower for subjective criteria, including weak effort (k=0.19) and cough in the first second (k=0.36). No agreement was observed for glottic closure (k=0). SR adjudication increased the proportion of acceptable tests from 54% to 65%. CONCLUSIONS: Reviewer variability significantly affects spirometric acceptability, with numerical criteria achieving higher agreement than subjective assessments. These discrepancies have implications for large-scale studies, data quality, and automated algorithm development. Cross-checking reviews is recommended to enhance reliability. Further standardization of acceptability criteria is warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy and reproducibility of a large language model in assessing spirometry test acceptability and interpretation: a performance study.Frontiers in medicine · 2026
- New software for assessment and interpretation of pulmonary function tests in adults: a comparative validation study.Frontiers in physiology · 2026
- Evaluation of an augmented reality-assisted system for standardizing spirometry (AR-SPIRO trial): Protocol for three-stage system development, agreement validation, and pragmatic implementation.Digital health · 2026
- Assessing inhalation technique: a comparison of three methods in patients with asthma and COPDnpj Primary Care Respiratory Medicine · 2026
- The impact of peak expiratory flow variability on FEV1 repeatabilityDiscover Respiratory Health · 2026
- Diagnostic uncertainty in respiratory sarcopenia: accuracy of available tests and clinical implications: a systematic review and meta-analysis.European geriatric medicine · 2026