Designing and fabrication of 4H-SiC neutron sensors for boron neutron capture therapy

Vu Thi Ha, Tatsuya Meguro, Kazutoshi Kojima, Hiroki Tanaka, Shin-Ichiro Kuroki

Key points

From the paper's conclusion
  • The results, which were obtained under the HWNIF condition with a high neutron flux in the range of 1 x 109 n/cm2.s and a cycle of 1.0 s, demonstrated the neutron detection capability of the 4H-SiC neutron sensor prototypes, despite signs of degradation due to neutron-induced damage and the TID effects.

Abstract

In this work, 4H-SiC neutron sensors for BNCT were designed, fabricated and tested under UV light and neutron irradiation. A pinned photodiode (PPD) structure and a parasitic bipolar junction transistor were designed for neutron sensor region. A large neutron conversion layer of 500 µm x 500 µm was formed on the PPD area by 10B ions implantation into a 700-nm Al layer. The output characteristics under UV light test have demonstrated the functional operation of the 4T-APS structure of SiC neutron sensors. No significant degradation was observed under E-3 tube with total neutron fluence of 1.8 x 109 n/cm2. The results, which were obtained under the HWNIF condition with a high neutron flux in the range of 1 x 109 n/cm2.s and a cycle of 1.0 s, demonstrated the neutron detection capability of the 4H-SiC neutron sensor prototypes, despite signs of degradation due to neutron-induced damage and the TID effects.

Read the paper