Relative Contact Load and PPG Modulation in a Low-LED-Drive Reflective Platform: A Single-Participant Feasibility Study
Yang Shang, Yucheng Huang, Ziang Li
Key points
From the paper's conclusion- These observations establish acquisition feasibility in the tested participant and arrangement, without establishing a causal pressure–response relationship, externally validated quality thresholds, or performance across users and devices.
Abstract
Reflective photoplethysmography (PPG) is sensitive to sensor–skin contact conditions. This engineering feasibility study examined synchronized relative-load voltage and Red/IR PPG in one participant using a MAX30101 sensor, nRF52833 microcontroller, and adjacent force-sensitive resistor (FSR). Five repeated trials followed a fixed low–medium–firm–medium-return sequence. Mean FSR voltages were 500.0 ± 22.1, 1179.8 ± 87.6, and 1996.6 ± 71.3 mV for the ascending stages. The 15 trial-stage summaries were repeated observations from the same participant. Higher FSR voltage was associated with lower IR AC/DC (within-trial r = −0.886; descriptive trial-fixed-effect slope, −0.0223 percentage points per 100 mV). Mean IR AC/DC was 0.560 ± 0.152% at low load, 0.218 ± 0.011% at firm load, and 0.381 ± 0.099% during medium return, compared with 0.330 ± 0.074% at the initial medium stage. Preliminary artifact controls documented setup-specific waveform changes. The adjacent FSR provides an uncalibrated spatial proxy rather than optical-site force or pressure. Fixed loading order prevents separation of load from time, adaptation, and loading history. These observations establish acquisition feasibility in the tested participant and arrangement, without establishing a causal pressure–response relationship, externally validated quality thresholds, or performance across users and devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Mechanical Contact Conditions in Wearable Reflectance Photoplethysmography: Scoping Review.JMIR mHealth and uHealth · 2026
- A Center-of-Pressure Guided Finger-Press Sensor for Cuffless Blood Pressure Estimation.Sensors (Basel, Switzerland) · 2026
- Reliable wrist PPG monitoring by mitigating poor skin sensor contactScientific Reports · 2025
- Accuracy of Photoplethysmography-Derived Pulse Rate Variability Compared with Electrocardiography-Derived Heart Rate Variability: A Systematic Review and Meta-Analysis.Sensors (Basel, Switzerland) · 2026
- Signal Quality of Reflective-Mode Photoplethysmograms Across Anatomical SitesSensors · 2026
- Performance validation of a new finger-root-adapted ring-shaped oximeter sensor during stepwise steady-state inspired hypoxia.Journal of clinical monitoring and computing · 2026