Machine Learning Field Calibration of a Low-Cost Carbon Nanotube Sensor Array Utilizing Near-Road Ambient Ozone Measurements.
Gautam Naishadham, Elena Bekyarova, Krishna Naishadham
Wi-Sense, Inc., Atlanta, GA 30082, USA. · Carbon Solutions (United States), Riverside, CA 92507, USA.
Key points
From the paper's conclusion- Prediction on an unseen data block (not used in training and cross-validation) reveals that the array discriminates ambient ozone levels above 20 ppb with RMSE < 4 ppb, and R2 = 0.9.
Abstract
Real-time monitoring of environmental pollutants, such as ozone and particulate matter, using low-cost sensors, has become increasingly important in various fields, including wearable devices, public health and occupational safety. Nanotechnology-based sensors offer significant advantages over commercial metal oxide semiconductor and electrochemical sensors, including low-temperature operation, low power consumption, enhanced sensitivity, and chemical functionalization of nanomaterials for selective gas detection. In this study, we integrate carbon nanotube (CNT) sensors based on thin films of three novel functionalized nanomaterials into a compact five-element sensor array and conduct ambient ozone measurements at a near-road location in downtown Atlanta, an area characterized by high vehicular emissions. Three distinct ozone-sensitive functionalized CNTs are investigated, namely, (a) Pt-CNT, comprising pristine CNTs decorated with platinum nanoparticles (~2 nm in diameter); (b) PBD-CNT, a composite of pristine CNTs and polybutadiene (PBD); and (c) ODA-CNT, comprising pristine CNTs covalently modified with octadecylamine (ODA) functional groups. The ambient measurements are performed during summer when the ozone levels are high due to abundant sunshine and high concentrations of precursor pollutants emitted by vehicular traffic. The field data comprising impedance magnitude and phase responses of the five sensors are analyzed using machine learning regression employing ensembles of gradient-boosted and random forest decision trees with blocked cross-validation. The sensor array is collocated with a precision ozone monitor, which provides the reference ambient ozone concentrations to calibrate the CNT sensor elements. Prediction on an unseen data block (not used in training and cross-validation) reveals that the array discriminates ambient ozone levels above 20 ppb with RMSE < 4 ppb, and R2 = 0.9.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Machine Learning Prediction of Analyte-Induced Fluorescence Perturbations in DNA-Functionalized Carbon NanotubesNano Letters · 2026
- Hybrid Deep Learning Neural Networks for Small-Molecule Organic Amine Gas Recognition and Prediction Based on MOS Sensor Array.Sensors (Basel, Switzerland) · 2026
- Enhanced Room Temperature NO2 Detection by Carbon Nanofibers and Single-Walled Carbon Nanotubes: Experimental and Molecular DynamicsChemosensors · 2025
- Carbon Nanotube-Based Gas Sensors: Sensing Mechanisms, Functional Interfaces, Gas-Specific Performance, and Flexible/Wearable Integration.Sensors (Basel, Switzerland) · 2026
- RBFNN-Based PPy/GaN sensor array with wide dynamic range and sub-ppb detection for accurate ammonia identification in human exhaled breathMicrosystems & Nanoengineering · 2026
- Physics-aware calibration neural network for automated TCAD-to-experiment modeling of carbon nanotube field-effect transistorsMoore and More · 2026