Sensors (Basel, Switzerland)

Validity Assessment of Wrist-Worn Electrodermal Activity Measurement During Relaxation and Acute Stressor Tasks: Comparing the EmbracePlus and Shimmer3 GSR.

Manon W H Peeters, Bo A C Dirks, Sjoerd Kesteloo, Bart van Boxtel, Petra Heck, Gerard Schouten

School for Allied Health Professions, Fontys University of Applied Sciences, Postbus 347, 5600 AH Eindhoven, The Netherlands. · School for Information & Communication Technology, Fontys University of Applied Sciences, Postbus 347, 5600 AH Eindhoven, The Netherlands.

Key points

From the paper's conclusion
  • This study evaluated the newly introduced EmbracePlus against the Shimmer3 GSR+, used as a research-grade wrist-based comparator.

Abstract

Wearable electrodermal activity (EDA) monitoring enables assessment of stress-related physiological changes in daily life, but new devices require validation. This study evaluated the newly introduced EmbracePlus against the Shimmer3 GSR+, used as a research-grade wrist-based comparator. Healthy volunteers (n = 53; n = 41 valid paired recordings) completed breathing exercises, a Stroop task, and a virtual reality (VR) height stressor while wearing both devices simultaneously on the non-dominant wrist. Agreement was assessed at signal, parameter, and event levels for mean phasic activity, skin conductance level (SCL), summed skin conductance response (SCR) amplitude/min, and SCR count/min. Mean signal-level cross-correlation was r = 0.30. At the parameter level, only 44% of SCL, 24% of SCR amplitude/min, and 5% of SCR count/min differences fell within a priori acceptable boundaries, and regression analyses indicated proportional bias for mean phasic activity, summed SCR amplitude/min and SCR count/min. At the event level, Shimmer3 detected increases across all four EDA metrics during both stress tasks (all p < 0.001). EmbracePlus showed increases across all four metrics during VR (all p < 0.001), whereas during Stroop only mean phasic activity (p = 0.001) and SCL (p < 0.001) increased. EmbracePlus and Shimmer3 are therefore not interchangeable for absolute EDA assessment, and EmbracePlus validity depends on stressor intensity and metric selection.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Validity Assessment of Wrist-Worn Electrodermal Activity Measurement During Relaxation and Acute Stressor Tasks: Comparing the EmbracePlus and Shimmer3 GSR — Science Skim