Dual X-Ray and Optical System for In Situ Monitoring of Solids Fraction in Tailings.
Talwinder Kaur Sraw, Bo Yu, Xiaoxuan Liu, Jason Ng, Andrea Sedgwick, Manisha Gupta, Robert Fedosejevs, Ying Yin Tsui
Department of Electrical and Computer Engineering, University of Alberta, 9211 116 Street NW, Edmonton, AB T6G 1H9, Canada. · Centre for Energy and Environmental Sustainability, Northern Alberta Institute of Technology, 10210 Princess Elizabeth Ave NW, Edmonton, AB T5G 0Y2, Canada.
Key points
From the paper's conclusion- A sensing apparatus was developed to quantify solids fraction within a 3 m test column, demonstrating real-time monitoring of tailings sedimentation.
Abstract
A sensing apparatus was developed to quantify solids fraction within a 3 m test column, demonstrating real-time monitoring of tailings sedimentation. This integrated system employs a low-power infrared laser coupled with scattering detectors alongside a low-activity radioactive source and detector. Both methodologies utilize non-destructive principles to evaluate the concentration of solids. Specifically, X-ray transmission through the medium provides a measure of the solids fraction, as X-ray attenuation corresponds to the solids percentage. Concurrently, the optical component detects radiation backscatter from the sample. To ensure precision, both instruments underwent calibration using reference materials. Experimental evaluations were conducted using two configurations: a fixed array and a depth-profiling setup. The former utilized a 2.5 m plastic column with integrated sensors placed within a settling vessel to acquire data at discrete depths. Conversely, the depth-profiling configuration involved a submersible mobile unit lowered into the column to characterize the solids profile at arbitrary depths.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Solid–Liquid Impurity Separation in Shale Gas Gathering and Transportation Systems: Method Development and Field ApplicationEnergy & Fuels · 2026
- Investigation and correction of size-dependent transport efficiencies of microparticles in SP ICP-MS analysisTalanta · 2026
- Quantitative assessment of the size and concentration measurement in a novel flow cytometer.The Review of scientific instruments · 2026
- Development of a 3D-Printed Two-Stage Virtual Impactor for Radioactive Aerosol Size Classification and Direct Analysis.ACS omega · 2026
- Predicting field calibration factors for suspended sediment monitoring based on particle size and shapeJournal of Environmental Management · 2026
- Cumulative Spectroscopic Detection for Taylor Dispersion Analysis of NanoparticlesAnalytical Chemistry · 2026