SNR-Weighted Layer-Conditioned Magnetometer Array with Ellipsoid Calibration and Neural Network Initialization for Capsule Endoscopy Localization Under Asymmetric Sensor Visibility.
Omid Yaghoobian, Khan A Wahid
Department of Electrical and Computer Engineering, University of Saskatchewan, 57 Campus Drive, Saskatoon, SK S7N 5A2, Canada.
Key points
From the paper's conclusion- We present a two-layer magnetometer array with three algorithmic contributions: per-sensor SNR gating excluding saturated or low-SNR sensors before optimization; layer-conditioned Levenberg-Marquardt estimation on a five-dimensional manifold via spherical parameterization; and SNR-weighted fusion with geodesic interpolation on S2, combined with layer-wise ellipsoid calibration and neural network i
Abstract
Passive magnetic localization for wireless capsule endoscopy degrades under involuntary gastrointestinal motion: displacement toward one sensor layer simultaneously saturates near-side magnetometers while driving far-side sensors below the noise floor, a failure mode we term asymmetric layer visibility. Existing single-layer and multi-layer systems do not model this explicitly, and full-array Jacobian conditioning degrades progressively with displacement even where a layer-conditioned decomposition remains comparatively stable. We present a two-layer magnetometer array with three algorithmic contributions: per-sensor SNR gating excluding saturated or low-SNR sensors before optimization; layer-conditioned Levenberg-Marquardt estimation on a five-dimensional manifold via spherical parameterization; and SNR-weighted fusion with geodesic interpolation on S2, combined with layer-wise ellipsoid calibration and neural network initialization. Eight LIS3MDL tri-axial magnetometers in a 15×15×30 cm3 two-layer array were validated across 150 Monte Carlo trials using a 4-DOF robotic arm. Mean positional error was 2.32±0.35 mm and orientational error 2.15±0.88° at 12.4 Hz, with 93.85% convergence success under displacement-induced asymmetric visibility events of 1.00-4.00 cm-over 70% reduction over single-layer and layer-agnostic baselines. Hardware generalization was confirmed across three magnetometer types. The SNR weight ratio also yields a passive discriminant between capsule displacement and global array motion, cross-validated AUC 0.924±0.045, requiring no additional hardware. Reported accuracy is specific to the LIS3MDL magnetometer used in this study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Electromagnetic Localization 6-DoF Pose Estimation with TGNet-Guided Initial Value RefinementMathematics · 2026
- A Two-Element Capsule Antenna for Near-Field Polarization MIMO in Wireless Capsule EndoscopyarXiv · 2026
- On the Pose Estimation of an Untethered Microrobot via Magnetic Localization and Optical Tracking.Micromachines · 2026
- Oil-chamber profile optimization design for drag reduction in an underactuated capsule robot with a suspended rotorJournal of Physics: Conference Series · 2026
- Deep self-organizing competitive neural network for magnetically controlled capsule endoscopy-based auxiliary diagnosis of clinically significant gastric lesionsWorld Journal of Clinical Cases · 2026
- Anatomically-Based Biomagnetic Source Imaging of Gastric Slow WavesBioengineering · 2026