Microwave Radar Sensing for Non-Invasive Intra-Abdominal Pressure Monitoring: A Simulation-Based Analysis with Phantom Testing.
Salar Tayebi, Ashkan Zarghami, Cheng Chen, Wojciech Dabrowski, Manu L N G Malbrain, Johan Stiens
Department of Electronics and Informatics, Vrije Universiteit Brussel, 1050 Brussel, Belgium. · First Department of Anaesthesiology and Intensive Therapy, Medical University of Lublin, 20-090 Lublin, Poland.
Key points
From the paper's conclusion- The present study should be considered a preliminary proof of concept. Clinically, the technology is currently more suitable for early warning and trend monitoring than for precise absolute IAP measurement, and it does not yet replace standard intravesical measurements. Its ability to support clinical decision-making, including guiding fluid therapy, requires prospective validation in patients.
Abstract
BACKGROUND: Intra-abdominal pressure (IAP) has recently been recognized as a new vital sign in critically ill patients. Microwave reflectometry has been proposed as a potential approach for non-invasive IAP measurement. However, systematic investigation on how individual anatomical and geometric factors influence changes in the microwave reflection response of the abdominal compartment is limited. Complementary information regarding illumination frequency and specific absorption rate (SAR) also warrants consideration. OBJECTIVE: This study aimed to advance the current knowledge on using microwave radar-based sensors in IAP monitoring by studying the most influencing factors. The penetration depth and spot size versus radiation frequency is studied as well. Information on energy deposition due to radio-frequency exposure is investigated too. METHODS: Numerical simulations were performed using abdominal models adjusted to represent different IAP levels. Reflection signal features were analyzed in relation to IAP-induced changes, and SAR was calculated using human models. Subsequently, a radar sensor prototype was tested on a benchtop abdominal phantom. Lin's concordance correlation analysis was used to evaluate absolute agreement between radar-estimated IAP and reference IAP. Additional statistical analyses assessed bias, precision, concordance, and risk levels. RESULTS: Sagittal abdominal diameter was the dominant factor affecting the microwave reflection response. Reflection amplitude showed a periodic trend consistent with abdominal displacement corresponding to multiples of half-wavelength values of the applied electromagnetic waves. Numerical SAR simulations showed increasing SAR with frequency while remaining below the applicable exposure limits under the investigated conditions. The radar sensor showed a bias of 0.43 mmHg and a precision of 2.55 mmHg. Concordance analysis among the paired changes remaining after application of the predefined exclusion criteria showed agreement in the direction of IAP change. CONCLUSION: The present study should be considered a preliminary proof of concept. Clinically, the technology is currently more suitable for early warning and trend monitoring than for precise absolute IAP measurement, and it does not yet replace standard intravesical measurements. Its ability to support clinical decision-making, including guiding fluid therapy, requires prospective validation in patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Positioning-robust broadband microwave transmission as a minimal-complexity approach to thoracic dielectric sensing: simulation and phantom analysis.Scientific reports · 2026
- Noninvasive and Quantitative Intracranial Pressure Estimation Using Microwave TechniqueIEEE Sensors Journal · 2025
- A Wearable Open-Domain Bioimpedance Sensor for Noninvasive Assessment of Deep Abdominal Fat Distribution.ACS sensors · 2026
- Characteristics of Abdominal Movement Signals Measured by a Wireless Abdomen-Worn Sensor During Home Sleep Apnea Testing: Pilot Observational Study.JMIR formative research · 2026
- Dielectric characterisation of human adrenal glands and adrenal tumours for the development of microwave ablation technologies for hypertension treatment.ArXiv · 2026
- Clinical validation of an intra-abdominal pressure monitoring system by comparison with pneumoperitoneum pressure.Journal of clinical monitoring and computing · 2026