Prognostic Value of Traditional and Optimised Vital Sign Thresholds for Risk of Serious Adverse Events in Continuously Monitored Hospitalised Patients.
Nikolaj Aagaard, Markus Harboe Olsen, Jesper Mølgaard, Katja Kjær Grønbæk, Mikkel Elvekjær, Eske K Aasvang, Christian S Meyhoff
Department of Anaesthesia and Intensive Care, Copenhagen University Hospital-Bispebjerg and Frederiksberg, 2400 Copenhagen, Denmark. · Department of Anaesthesia, Centre for Cancer and Organ Diseases, Copenhagen University Hospital-Rigshospitalet, 2100 Copenhagen, Denmark.
Key points
From the paper's conclusion- This study aimed to assess the prognostic ability of combinations of threshold severity and duration in continuous vital sign monitoring for SAEs in acute medical and surgical patients.
Abstract
Technological advances enable continuous vital sign monitoring with wireless biosensors in general wards. However, traditional thresholds used in routine monitoring do not effectively identify serious adverse events (SAEs). This study aimed to assess the prognostic ability of combinations of threshold severity and duration in continuous vital sign monitoring for SAEs in acute medical and surgical patients. Data on heart rate (HR), respiratory rate (RR), and peripheral oxygen saturation (SpO2) were collected from four prospective observational studies and two randomised controlled trials. The primary outcome was an SAE within 30 days. Traditional thresholds were based on the National Early Warning Score, and optimised thresholds were derived by identifying threshold values and durations with the highest area under the receiver operating characteristic curve (AUROC) for any SAE and specific SAE types. Of 1492 patients, 1077 were included in HR and RR analyses and 761 in SpO2 analyses; 30% experienced an SAE. For any SAE, traditional thresholds had AUROC values of 0.47-0.50 (95% CI 0.43-0.54). Optimised thresholds also showed limited ability (AUROC 0.51, 95% CI 0.44-0.56). In contrast, 18 thresholds for specific SAE types had AUROC > 0.70; the highest AUROC was observed for HR > 104 beats/min for ≥36 min for upcoming respiratory SAEs (AUROC 0.82, 95% CI 0.74-0.90). Optimised thresholds showed low prognostic ability for any SAE but higher AUROC than traditional thresholds. Several thresholds demonstrated moderate prognostic ability for specific SAE groups and warrant validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Artificial Intelligence-Driven Continuous Biosignal Analysis in Anesthesia and Critical Care: Detection, State Assessment, and Adverse-Event Prediction: A Critical Narrative Review.Bioengineering (Basel, Switzerland) · 2026
- Artificial Intelligence-Enhanced Early Warning Systems for Early Detection of Clinical Deterioration in Hospitalized Patients: Current Clinical Applications and Future Directions.Maedica · 2026
- Continuous versus intermittent, non-invasive monitoring of vital signs in hospitalised, ward based adult patients: A systematic review and narrative synthesis.Journal of the Intensive Care Society · 2026
- Predictive performance of vital sign–based prediction models for in-hospital cardiac arrest: a systematic review and meta-analysisBMC Cardiovascular Disorders · 2026
- Development and validation of a clinical wearable deep learning based continuous inhospital deterioration prediction modelNature Communications · 2025
- Prognostic value of disease severity and mechanical ventilation intensity in acute respiratory distress syndrome: analysis of the LUNG SAFE cohortEuropean Respiratory Journal · 2025