Impact of Regeneration on the Dosimetric Properties of LiF:Mg,Cu,P (MCP-N) Detectors.
Patrycja Gordon, Katarzyna Matusiak, Mariusz Kłosowski, Aleksandra Jung
Maria Skłodowska-Curie National Research Institute of Oncology, Kracow Branch, 30-059 Krakow, Poland. · AGH University of Krakow, Faculty of Physics and Applied Computer Science, Department of Medical Physics and Biophysics, 30-059 Krakow, Poland.
Key points
From the paper's conclusion- Overall, regenerated detectors, despite reduced sensitivity, achieve performance comparable to standard detectors after stabilization.
Abstract
This study evaluates the dosimetric properties of standard (SD) and regenerated (RD) MCP-N thermoluminescent detectors over a wide dose range from mGy to Gy, with particular emphasis on the influence of readout mode, dose level, and temporal stability on detector performance. A total of 17 SD and 18 RD detectors were analyzed under standard conditions using two readout modes: a three-step READER mode and a continuous ANALYSER mode. Both detector types exhibit good linearity; however, deviations from ideal behavior at higher doses suggest the involvement of additional trapping and recombination processes. Glow-curve analysis indicates modifications in RD detectors, reflected by enhanced low-temperature peaks and slight shifts in peak position. Temporal effects also influence detector response, with stabilization observed after repeated irradiation-readout cycles. Fading remains low, with relatively stable signal levels over time. The results show that the ANALYSER mode provides improved stability and reproducibility compared to the READER mode, especially at higher doses, where increased variability and certain limitations were observed. Overall, regenerated detectors, despite reduced sensitivity, achieve performance comparable to standard detectors after stabilization. Optimization of readout conditions, particularly the use of ANALYSER mode, is essential for ensuring reliable dosimetric performance across a wide dose range.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Thermally Assisted OSL in Sedimentary Quartz: Two-Stage Kinetics, Extended Dose Response and a Modified SAR Baseline.Luminescence : the journal of biological and chemical luminescence · 2026
- Advancing understanding of temporal performance and behaviour in consumer-grade electronic radon monitors.Radiation protection dosimetry · 2026
- Thermoluminescence of undoped Na2Gd2B2O7: An energetically heterogeneous trap manifold and evidence for non-monotonic post-irradiation storage behaviour.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Engineering trap dynamics in Ba2ZnSi2O7 phosphors: a study on rare-earth co-doping for efficient TL dosimetry.RSC advances · 2026
- Experimental validation of the scintiPulses shot-noise model for coincidence resolving-time dependencies in liquid scintillation counting.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Validation of a detector-response correction model for in vivo dosimetry in high-dose-rate prostate brachytherapy.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026