High-Frequency Weak Magnetic-Field Detection and Dose Calibration for Micro-Magnetic Stimulation: A Review.
Lei Tian, Hang Tu, Jiahao Fei, Shuai Cao, Yongxiang Gao, Yameng Zhou, Jianpeng Song
School of Control Science and Engineering, Tiangong University, Tianjin 300387, China. · School of Life Sciences, Tiangong University, Tianjin 300387, China.
Key points
From the paper's conclusion- Overall, this review provides a structured overview of micro-coil magnetic-field measurement, stimulation dose standardization, and the quantitative frameworks for closed-loop neuromodulation microsystems.
Abstract
High-frequency weak magnetic fields generated by submillimeter micro-coils are key physical parameters associated with stimulation dose in micro-magnetic stimulation (μMS). However, their low amplitude, highly localized near-field distribution, frequency-dependent probe response, and sensitivity to coupling variations make accurate measurement and dose calibration challenging. This review summarizes recent advances in high-frequency weak magnetic-field detection techniques for micro-coil systems. It discusses key measurement challenges, including probe parasitics, impedance mismatch, phase distortion, spatial averaging, and dynamic detuning. Existing methods are classified into direct magnetic-field measurement, resonance-enhanced detection and calibration, spatial-field mapping reconstruction, and adaptive compensation. These methods are compared across five dimensions: frequency range, sensitivity, spatial resolution, system complexity, and online integration capability. The review further highlights the need to establish reliable and traceable relationships among detector outputs, local magnetic fields, and stimulation dose. Future directions include broadband adaptive measurement, online dose calibration, multichannel magnetic-field imaging, multichannel magnetic-field component decoupling, and physics-constrained data-driven field estimation. Overall, this review provides a structured overview of micro-coil magnetic-field measurement, stimulation dose standardization, and the quantitative frameworks for closed-loop neuromodulation microsystems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Micro-magnetic stimulation for neuromodulation: technological evolution from discrete inductors to MEMS and wireless passive systems.Medical engineering & physics · 2026
- High-Resolution 3D Micro-CT Database of Transcranial Magnetic Stimulation (TMS) Coils for More Accurate Modelling, Technology Benchmarking, and Progress in Coil DesignarXiv · 2026
- Low Intensity Multi-Channel Steering TMS Array for Network Level Neuromodulation.bioRxiv : the preprint server for biology · 2026
- Development of RF Probes and Preamplifiers for MRI at Low FieldsNMR in Biomedicine · 2026
- Dictionary-based reconstruction of spatio-temporal 3D magnetic field images from a quantum diamond microscopeOptica · 2026
- High‐Density Magnetomyography With Optical Magnetometers Reveals Spatiotemporal Patterns of Evoked Muscle ActivityAdvanced Materials Technologies · 2026