Processes

Quality Engineering Assessment of Laboratory Workflows for Organochlorine Pesticide Monitoring in Agricultural Soils: A Documentary Case Study

Oana Roxana Chivu, Daniela Cîrțînă, Mihaela Tamara Leonte, Liviu Marius Cîrțînă

Key points

From the paper's conclusion
  • A third report from the same reporting laboratory, covering 373 pesticide entries in a potato-tuber sample, served as a reference for entry-level method, reporting-limit and accreditation labelling: 14.6% of the 335 direct determinations were marked outside the accreditation scope.

Abstract

This exploratory documentary study examines the sampling-to-reporting workflow behind two third-party soil test reports on organochlorine pesticides: an in-house analysis of a sample from Bulbucata (Giurgiu County), with a quantified residue profile, and a subcontracted analysis of a sample from Găești (Dâmbovița County), with all analytes reported below the 0.01 mg/kg determination limit. Four quality-engineering tools were applied to the documents only: a file-based review against ISO/IEC 17025:2017, organised as an evidence matrix; turnaround-time indicators derived from report metadata; a failure mode and effects analysis (FMEA) with explicit scoring criteria and a sensitivity analysis; and an individuals control chart framework, presented as formulae and a data-collection template for future routine monitoring. Receipt-to-report turnaround was 8 days for Bulbucata and 61 days for Găești; the declared analysis windows covered 75.0% and 13.1% of these intervals, and the report metadata locate 41 of the 61 Găești days between the end of the analysis window and the issuance of the subcontractor’s report, without identifying the activities performed. Neither report states a measurement uncertainty, and the Bulbucata bulletin carries no accreditation statement for the test, whereas the Găești report discloses the subcontractor and its accreditation. The FMEA ranked cross-contamination at manual sampling (RPN = 192), insufficient extraction recovery (160) and sorption on plastic packaging (147) highest; sampling remained first in 57 of 60 one-at-a-time ±1 scoring scenarios, whereas ranks 2–9 were unstable. A third report from the same reporting laboratory, covering 373 pesticide entries in a potato-tuber sample, served as a reference for entry-level method, reporting-limit and accreditation labelling: 14.6% of the 335 direct determinations were marked outside the accreditation scope. The study provides a reusable source-document table and evidence matrix for file-based reviews and identifies the operational records needed before laboratory performance can be assessed. All information is labelled by evidence level as reported (transcribed from the documents), derived (calculated from reported dates and counts), assessed (author judgement) or proposed.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Quality Engineering Assessment of Laboratory Workflows for Organochlorine Pesticide Monitoring in Agricultural Soils: A Documentary Case Study — Science Skim