Impact Attenuation Performance of a Lightweight, Ventilated Two-Zone Natural Rubber Hip Protector: Experimental Testing and Finite Element Optimization
Theerawat Petdee, Wiriya Thongruang, Satta Srewaradachpisal, Charoenyutr Dechwayukul, Boonsin Tangtrakulwanich, Varah Yuenyongviwat
Key points
From the paper's conclusion- This study used finite element (FE) modeling and drop-tower testing to develop a lightweight two-zone natural rubber hip protector (TZNRHP), comprising a stiff solid-rubber inner zone over the greater trochanter and a closed-cell rubber-foam outer zone.
Abstract
Hip fractures caused by sideways falls impose a major health and economic burden on older adults, yet hip protectors are often rejected because they are thick, heavy, and uncomfortable. This study used finite element (FE) modeling and drop-tower testing to develop a lightweight two-zone natural rubber hip protector (TZNRHP), comprising a stiff solid-rubber inner zone over the greater trochanter and a closed-cell rubber-foam outer zone. A hyperelastic–viscoelastic FE model was validated against drop-tower tests on a surrogate hip (4.75 kg impactor, 0.52 m drop), and protectors 9–15 mm thick with six ventilation-hole patterns were evaluated (n = 5 repeated impacts; one-way ANOVA with Tukey’s test). The FE model predicted peak forces within 5.4% for all validation conditions. Thickness significantly affected peak force (p < 0.001, η2 = 0.99), and the 11 mm protector was the thinnest design significantly below the 3100 N fracture criterion, reducing the peak force from 6232 ± 160 N (unprotected) to 2785 ± 79 N (55.3% attenuation) while being 28.3% lighter than the original 15 mm design. For a female-specific threshold of 2800 N, the 12 mm design was required. Thirteen 5 mm ventilation holes (8.4% open area) raised the peak force by only 2.6% (2856 ± 73 N; not significantly different from the least-perforated pattern), and the 11 mm protector reduced the maximum femoral von Mises stress from 9.2 to 5.83 MPa. The optimized TZNRHP is a promising lightweight, ventilated hip protector for older adults.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Bioinspired auxetic metamaterial liners and sockets for transtibial prostheses: Energy absorption and stress redistributionJournal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · 2025
- Comparing oral bisphosphonates, denosumab, hip protector, and bioactive hydrogel augmentation strategies for hip fracture prevention using subject-specific finite element models.Bone · 2026
- Influence of the pelvic bone and gluteal zone over the stress state at the interaction between the prosthetic socket and residual limb for people with transfemoral amputation.Frontiers in bioengineering and biotechnology · 2026
- FallGown: A Soft Pneumatic Wearable Device for Fall PreventionarXiv · 2026
- Thresholds of Failure: Biomechanical Prediction of Hip and Acetabular Fractures in High-Velocity Dashboard ImpactsApplied Sciences · 2026
- Clinical role, biomechanical aspects, and barriers to the use of hip protectors in fracture prevention: insights from an Italian Delphi consensus panel.Reviews in endocrine & metabolic disorders · 2026