Polymers

Finite Element Investigation of the Influence of Strut Diameter on the Mechanical Performance of Balloon-Expandable Biodegradable PLA/PDO Coronary Stents.

Elhadj Besseghier, Fatima Zohra Kettaf, Ahmed Ouadah Bouakkaz, Abdelkader Djebli, Ali Benhamena, Dursun Murat Sekban, Ecren Uzun Yaylacı, Merve Terzi,

Mechanical Engineering Department, Faculty of Technology, Hassiba Benbouali University of Chlef, Hay Salem, National Road N19-02000, Chlef 02000, Algeria. · Faculty of Mechanical Engineering, University of Science and Technology of Oran Mohamed-Boudiaf USTO-MB, Oran 31000, Algeria.

Key points

From the paper's conclusion
  • This study numerically investigates the influence of strut diameter on the deployment behavior of a balloon-expandable biodegradable stent with rhombic cell architecture using finite element analysis.

Abstract

This study numerically investigates the influence of strut diameter on the deployment behavior of a balloon-expandable biodegradable stent with rhombic cell architecture using finite element analysis. The stent material was represented by a 60/40 poly(lactic acid)/polydioxanone (PLA/PDO) blend. Four stent configurations with strut diameters of 0.15, 0.25, 0.35, and 0.50 mm were analyzed under identical deployment conditions. The numerical evaluation considered von Mises stress together with five deployment indicators: diametral strain, elastic recoil, dog-boning, foreshortening, and longitudinal retraction. The results show that increasing the strut diameter reduces elastic recoil, foreshortening, and longitudinal retraction, thereby enhancing post deployment dimensional stability. However, thicker struts also increase the dog-boning effect, indicating less uniform radial expansion. Among the investigated designs, the stent with a 0.35 mm strut diameter showed a balanced response between deployment uniformity and post deployment mechanical stability under the adopted numerical assumptions. For this configuration, elastic recoil, foreshortening, longitudinal retraction, and dog-boning were approximately 6.2%, 10%, 37%, and 8.9%, respectively. These findings provide practical design guidance for biodegradable polymeric vascular stents.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Finite Element Investigation of the Influence of Strut Diameter on the Mechanical Performance of Balloon-Expandable Biodegradable PLA/PDO Coronary Stents — Science Skim